Chapter 4 Operating instructions
FLUSHING PUMP OFP-2
37
(i) Remove the MAJ-1606 instrument channel adaptor from the endoscope.
(ii) Remove the MAJ-1607 instrument channel tube from the water container. Drain the
MAJ-1607 instrument channel tube by pressing the footswitch to supply air.
WARNING
The expelled liquid contains infectious material. Handle it appropriately.
(iii) Press the standby switch to "standby mode". If finishing for the day, press the power
switch to turn off the power supply.
(iv) Open the pump head lever and remove the MAJ-1607 instrument channel tube from the
pump head.
(v) Discard the MAJ-1606 instrument channel adaptor and the MAJ-1607 instrument
channel tube in an appropriate manner.
12. Remove the MAJ-1681 or MAJ-1682 accessory port tube as follows:
(i) Disconnect the tubing from the EndoTherapy device.
(ii) Remove the tubing from the water container or disconnect spike from the saline bag.
Drain the tubing by pressing the footswitch to supply air.
WARNING
The expelled liquid contains infectious material. Handle it appropriately.
(iii) Press the standby switch to "standby mode". If finishing for the day, press the power
switch to turn off the power supply.
(iv) Open the pump head lever and remove the tube from the pump head.
(v) Discard the tubing in an appropriate manner.
13. If used, the water container and cap should be removed from the OFP-2 and cleaned and
sterilized prior to reuse.