Important Information — Please Read Before Use
FLUSHING PUMP OFP-2
12
• Prior to each use, ensure the tube connections are secure, watertight
and routed to eliminate potential flow restrictions caused by, but not
limited to, tube kinking or accidental tube compression. Keep the tubing
clear of the patient at all times. In the event of entanglement, stop the
procedure and re-route the tubing.
• Medical electrical equipment needs special precautions regarding EMC
and needs to be installed and put into service according to the EMC
information provided in their accompanying documents.
• It should be noted that portable and mobile RF communications
equipment can affect medical electrical equipment.
• The use of accessories and cables other than those specified may result
in increased emissions or decreased immunity of the medical electrical
equipment or system.
• Ensure that the OFP-2 is not used adjacent to or stacked with other
equipment (other than components of the OFP-2 or the endoscopy
system) to avoid electromagnetic interference.
• When the OFP-2 is used in close proximity to other equipment, the
medical electrical equipment or system should be observed to verify
normal operation in the configuration in which it will be used.