Chapter 4 Operating instructions
FLUSHING PUMP OFP-2
36
NOTE
• It may be found that the pump head moves from side to side during
operation. This is perfectly normal and is indicative of the pump head
attachment method.
• After the pump has stopped, the pump may rotate for approximately two
seconds in the reverse direction. After each flush, it may appear that
bubbles in the tube are flowing back to the water container from the
OFP-2 pump head. This is a perfectly normal operation to help reduce
the pressure within the tube.
7.
To refill the water container while using it on a patient, press the standby switch to put the
OFP-2 on standby, keep the pump head lever closed so the tube is not disturbed.
Pull the tube from the water container cap and remove the container from the unit. Unscrew
the cap, fill the container with the appropriate fluid then secure the cap. Place the container
back on the unit. Push the tube into the container cap up to the indicator line. Press the
standby switch to put the OFP-2 back into “run mode”.
CAUTION
• During the water container refilling procedure, ensure that the tube is
handled appropriately to prevent contamination.
• Should the tube become contaminated during the water container filling
procedure, replace with a new tube.
8.
After using it on a patient, remove the endoscope from the patient. If using this product to
clean the endoscope's auxiliary channel, please do so in accordance with the endoscope
instruction manual.
9.
If intending to use auxiliary channel flushing on the next patient, remove the MAJ-855
auxiliary tube or MAJ-1652 adaptor from the pump tube. On connection of the new
endoscope, new MAJ-855 auxiliary tube or MAJ-1652 adaptor must be connected to the
pump tube to the MAJ-1608 auxiliary channel tube, fill the auxiliary channel with water by
flushing from the endoscope.
10. Remove the MAJ-1608/MAJ-1652 auxiliary channel tube as follows:
(i) Remove the tube from the water container and drain the auxiliary channel tube by
pressing the footswitch to supply air.
WARNING
• The expelled liquid contains infectious material. Handle it appropriately.
• Do not supply air to the endoscope while it is inserted in the patient.
(ii) Press the power switch to turn off the power supply.
(iii) Disconnect the pump tube from the one-way valve (MAJ-855 or MAJ-1652).
(iv) Open the pump head lever, remove the tube from the pump head and dispose of it in an
appropriate manner.
(v) Isolate the pump from the mains electricity at the end of use.
(vi) Disconnection from mains supply is achieved by removal of mains lead from wall or
transformer outlet sockets.
11. Remove the MAJ-1606 instrument channel adaptor and the MAJ-1607 instrument channel
tube as follows: