Chapter 6 Care and Storage
FLUSHING PUMP OFP-2
44
Chapter 6 Care and Storage
6.1 Care
WARNING
• After wiping with a piece of moistened gauze, dry the OFP-2 thoroughly
before using it again. If it is used while still wet, there is the risk of an
electric shock.
• When cleaning the OFP-2, always wear appropriate personal protection
equipment. Blood, mucus and other potentially infectious material
adhering to the OFP-2 could pose an infection control risk.
CAUTION
• Do not use harsh or abrasive cleaning materials on the OFP-2 as
damage may result.
• Ensure all surfaces are thoroughly dried before reusing the OFP-2.
• Ensure footswitch is removed from the floor and stored safely to prevent
it being damaged.
• The pump has a good, broad chemical resistance to inorganic acids,
saline solutions, alkalis, some hydrocarbons and a large number of oils
and greases. It is suitable for wipe down but not for long term contact
with alcohols. The case may be damaged by contact with strong acids
or strong solvents.
• Do not clean the remote control socket and/or the mains power inlet.
Cleaning them can deform or corrode the contacts, causing damage to
the OFP-2.
• Do not autoclave or gas sterilize the OFP-2. These methods will
damage it.
• Do not wipe the external surface with hard or abrasive wiping material.
The surface will be scratched.
NOTE
• There are no user serviceable parts inside the pump head.
• If the pump head requires replacement or thorough cleaning, detach
and reattach the pump head using the guidelines given in Section
“Pump Head Replacement” on page 46.
If the OFP-2 is soiled, perform the following cleaning procedure immediately after use. If
cleaning is delayed, residual organic debris will begin to solidify, and it may be difficult to
effectively clean the OFP-2. The OFP-2 should also be cleaned routinely.
1.
Turn the OFP-2 OFF and disconnect the power cord.
2.
Should the equipment become soiled with blood or other potentially infectious materials, first
wipe off all gross debris using a neutral detergent, then wipe with a lint-free cloth moistened
with a surface disinfectant.
3.
Wipe the surface of the OFP-2 using a soft, lint-free cloth moistened with 70% ethyl or
isopropyl alcohol to remove dust, dirt, etc.
4.
Dry the OFP-2 with a clean, lint-free cloth.