Chapter 5 Cleaning and Sterilizing
FLUSHING PUMP OFP-2
43
WARNING
• The results of sterilization depend on various factors such as how the
sterilized instrument was packed or the positioning, method of placing
and loading of the instrument in the sterilization device. Please verify the
sterilization effects by using biological or chemical indicators. Also follow
the guidelines for sterilization issued by medical administrative
authorities, public organizations or the infection management sections
at each medical facility, as well as the instruction manual of the
sterilization device.
CAUTION
Sudden changes in temperature may damage the instruments.
Follow the guidelines of the medical facility when steam sterilizing. Operate the steam sterilizer in
accordance with the steam sterilizer instruction manual or with the maker instructions.
1.
Before steam sterilizing, the water container and cap must be thoroughly cleaned and dried.
Residual moisture inhibits sterilization.
2.
Place the water container and cap in a separate package, then place the sealed packages in
the steam sterilizer and steam sterilize them at the conditions given below.
For details on operation of the steam sterilizer, refer to the instruction manual for the steam
sterilizer or other manufacturer instructions.
CAUTION
Do not exceed a maximum autoclave temperature of 137°C (279°F), nor
an exposure time greater than 20 minutes, otherwise damage to the
equipment may result.
Minimum Hold Temperature
Minimum Exposure Time Guideline/Standard Reference
121 – 124
C
15 mins
EN 285: 2006, HTM 2010:1994
126 – 129
C
10 mins
EN 285: 2006, HTM 2010:1994
132
C
4 mins
ANSI/AAMI ST79:2006
134 – 137
C
3 mins
ANSI/AAMI ST79:2006
EN 285: 2006
HTM 2010:1994