Chapter 4 Operating instructions
FLUSHING PUMP OFP-2
33
Chapter 4 Operating instructions
WARNING
• To prevent inadvertent operation of the pump while fitting/changing the
tube, the pump head lever is fitted with a cutout which prevents the
pump head from operating when the pump head lever is opened.
• Each day before use, confirm that the pump head cutout is correctly
working by visually checking that the pump rollers stop when the pump
head lever is opened. If the cutout is found to be faulty, stop using the
OFP-2 and contact Olympus.
• Before inserting the endoscope into the patient's body, prime the fluid
path and check irrigation tubing and adapter for signs of leakage. If fluid
leaks from the irrigation tubing and/or adapter, check tightness of
connections. If fluid is found to still leak from the irrigation tubing and/or
adapter, replace the affected item before continuing with the procedure.
• MAJ-1681 and MAJ-1682 tube sets must be primed without the
EndoTherapy accessory attached so air can be removed. Flush the
pump at speed setting 5-9 until all air pockets are gone, then attach the
EndoTherapy accessory.
1.
Press the power switch on the OFP-2 front panel. The flow rate indicators sequence up and
down during startup and the standby switch and lowest setting flow rate indicator illuminates
green to indicate the unit is in “ON”.
WARNING
Check correct illumination of all indicators during the startup sequence.
If an indicator fails to illuminate, then stop using the OFP-2 and contact
your nearest Olympus service centre.
NOTE
• Pressing the standby switch at this stage will change the standby switch
indicator to amber, indicating that the unit is in “standby mode”. Pressing
the standby switch again will change the standby switch indicator back
to green, indicating that the unit is “run mode”. Any previously selected
flow rate will be stored in memory, even when in “standby mode”, until
the unit is switched OFF.
• Before opening the pump head lever for tube replacement/adjustment,
press the standby switch to put the unit into “standby mode”. On
subsequent closure of the pump head lever, press the standby switch
again to switch the unit back into “run mode”.
2.
Press the
flow rate control to maximize the water flow. Press and hold down the footswitch
until water exits from the endoscope or EndoTherapy device. Then release the footswitch to
stop the water flow.