Operation Manual
Troubles and Countermeasures
/ When Error Codes Are Displayed
5-5
Page
Error No.
Error details
Remarks
E0313
A Rotation Overheating of motor driver
A
Suction error
E0314
A1 No detection of the vacuum ON status within the specified time
U007
E0315
A1 No detection of the vacuum OFF status within the specified time
U007
E0316
A2 No detection of the vacuum ON status within the specified time
U008
E0317
A2 No detection of the vacuum OFF status within the specified time
U008
Orientation flat detection error
E0401
M
No orientation flat detection within the specified time
U013,U014,U021
E0402
M
Overheating of the motor driver
Back Macro rotary movement errors
(
for the LMB model only
)
E0501
L Rotation No sensor detection after movement to the home
position
U401
E0502
L Rotation No status change of the home position sensor within the
specified time
U401
E0503
L Rotation Not in place at the start of operation
U401
E0507
L Rotation Movement set to an impossible position
E0508
L Rotation No initialization within the specified time
U401
E0511
L Rotation Not in a position where it can be initialized
E0513
L Rotation Overheating of the motor driver
Elevator movement errors
E0601
E
No sensor detection after movement to the upper limit
U101
E0606
E
Not in place at the start of operation
E0607
E
Movement set to an impossible position
E0608
E
No initialization within the specified time
E0609
E
Motor step-out was detected during vertical movement.
U102
、
U103
E0610
E
The wafer designated to be inspected does not match the wafer
detection mapping.
E0612
E
The driver origin signal cannot be detected.
E0613
E
Overheating of the motor driver
Top Macro
vertical
movement errors
(
for the LMB model only
)
E0701
M Vertical No detection of the upper limit sensor within the specified
time
U301
E0702
M Vertical No detection of the lower limit sensor within the specified
time
U302
E0705
M Vertical Two or more sensors ON
E0706
M Vertical Not in place at the start of operation
E0707
M Vertical Movement set to an impossible position
E0708
M Vertical No initialization within the specified time
M302
E0709
M Vertical No detection of the vacuum ON status within the
specified time
U009
E0710
M Vertical No detection of the vacuum OFF status within the
specified time
U009
Содержание AL120-6 Series
Страница 2: ...This page intentionally left blank ...
Страница 9: ...Operation Maintenance Manual Introduction Safety Precautions i 7 Page AL120 86 Series Unit mm ...
Страница 16: ...Introduction Conformity Standards Operation Maintenance Manual i 14 Page ...
Страница 17: ...Operation Maintenance Manual Introduction Conformity Standards i 15 Page ...
Страница 18: ...Introduction Conformity Standards Operation Maintenance Manual i 16 Page This page intentionally left blank ...
Страница 21: ...Operation Manual Table of Contents C 3 Page 4 When a Wafer is on the Vacuum Stage 5 10 ...
Страница 22: ...Operation Manual Table of Contents C 4 This page intentionally left blank Page ...
Страница 58: ...Inspections Operation Manual 3 14 This page intentionally left blank Page ...
Страница 78: ...Troubles and Countermeasures Automatic Unloading Operation Manual 5 12 Page This page intentionally left blank ...
Страница 79: ...This page intentionally left blank ...