Operation Manual
Troubles and Countermeasures
/ When Error Codes Are Displayed
5-3
Page
5-2 When Error Codes Are Displayed
When an error occurs, the loader detects the error and stops all
operations. At the same time, a blinking error code appears on the liquid
crystal panel
and a buzzer sounds.
Press the [M1] menu button
to silence the buzzer.
When there are no wafers in the transfer path, turn the main switch OFF
and then turn it back to ON.
When there are wafers in the transfer path, turn the main switch OFF,
remove the wafer(s) using tweezers or similar tools, and ensure that the
cause is dealt with.
When an adsorption error has occurred, the loader may be unable to
hold wafers securely. Take care not to perform automatic unloading
without dealing with the cause. Wafers may be damaged.
If any similar errors occur during operation, please contact the OLYMPUS
distributor for your loader.
Please give the distributor the following information:
Product name:
Serial No.:
Error code:
Phenomenon:
E0101
OFF
Caution
Содержание AL120-6 Series
Страница 2: ...This page intentionally left blank ...
Страница 9: ...Operation Maintenance Manual Introduction Safety Precautions i 7 Page AL120 86 Series Unit mm ...
Страница 16: ...Introduction Conformity Standards Operation Maintenance Manual i 14 Page ...
Страница 17: ...Operation Maintenance Manual Introduction Conformity Standards i 15 Page ...
Страница 18: ...Introduction Conformity Standards Operation Maintenance Manual i 16 Page This page intentionally left blank ...
Страница 21: ...Operation Manual Table of Contents C 3 Page 4 When a Wafer is on the Vacuum Stage 5 10 ...
Страница 22: ...Operation Manual Table of Contents C 4 This page intentionally left blank Page ...
Страница 58: ...Inspections Operation Manual 3 14 This page intentionally left blank Page ...
Страница 78: ...Troubles and Countermeasures Automatic Unloading Operation Manual 5 12 Page This page intentionally left blank ...
Страница 79: ...This page intentionally left blank ...