If the Custom Calibration was performed with a different set point temperature than what you
plan on using (e.g Custom Calibration had set point of 37C but new experiment requires a set point
temperature of 39C).
2.
Use
Factory
settings optimized at 23°C Room Temperature. This is acceptable if your room
temperature is within 1°C from 23°C. However, for room temperature more than 1°C from 23°C this
may be not acceptable depending on the sensitivity of your sample.
3.
Choose to
Calibrate
the system in your lab at your Room Temperature, for maximum of accuracy
when using “Chamber Feedback
Mode”
. To do so follow the instructions on paragraph 7 (Chamber
Self-Calibration) and then press
the “Calibrate”
button; the automatic calibration procedure will
start.
Tip
►
Note please it may take 1-2 hours to complete the calibration.
Figure 15. Chamber calibration progress
Just wait until the following dialog box appears:
Figure 16. Chamber Calibration result
Figure 17 shows how to select the
Sample Feedback Mode
.