21
Figure 19. Incubator Closed
6.2.1.3
Objective Heater
Figure 20 shows how to enable/disable the Objective Heater (OBJ COLLAR ####). The Objective Heater is
disabled by default. Read this paragraph only if your system is equipped with Objective Heater (OBJ COLLAR
####).
Figure 20. Objective Heater enabling
Tip
►
Before enabling it, make sure it is already installed around your objective, and that it has been
correctly connected to the Temperature Control Unit.
When enabling and saving the Objective Heater the following Screenshot will appear:
Figure 21. Objective Heater Calibration Page
The temperature of the Objective Heater is factory calibrated at room temperature of 23°C, with a 3
degrees offset relative to the sample set-point temperature of 37°C. This offset value is appropriate for a
set point of 37°C. If your set point is different than 37°C the system will automatically apply the correct
1
2