Enable the Objective Heater as shown in Figure 20, the window in Figure 23 will open. Follow the
instructions on the message: first press
“CALIBRATE”
, next bring the immersion lens in contact with the
petri-dish.
Following this sequence of instructions is REQUIRED to place the Fine Gauge
Thermocouple in standby and act on the Objective Heater only
.
The calibration procedure typically takes longer than one hour. While calibrating an icon will appear on the
Homepage Temperature tab (see Figure 26) to remind you of the ongoing calibration. Once the calibration
is completed the icon will no longer appear.
In the Objective Heater Calibration progress window
press “Options” to access “Advanced Calibration
Parameters” (see
Figure 27). In this window you can adjust Accuracy and Standard Deviation, which will be
calculated over 10 minutes or readings. This allows setting more stringent parameters to the objective
heater calibration.
Tip
►
The higher are these values (therefore the less accurate and precise is the calibration), the less time
is taken by the calibration procedure to complete. Use default values to get the most accurate calibration.
When the calibration procedure is completed, the message shown in Figure 28 will appear.