31
6.2.2.5
Alarms
This page allows you to set deviation values (from set-point) beyond which you want the system to trigger
an Alarm.
Figure 44. How to enter in Alarms page
Tip
►
Flag
“Buzzer”
if you want the Alarm to be acoustical as well rather than just being displayed.
Figure 45. How to set the Deviation and Time Alarm
Figure 45. In this example the following settings are displayed (valid only after the system has reached its
stationary state): if Temperature Deviation from set point is 1°C or greater (i.e. if set point temperature is
37°C and the temperature reaches a value equal or less than 36.5°C or equal or more than 37.5°C) for a
period of time equal or longer than 15 minutes (“Temp time“ set in this example) then the system triggers
an Alarm.
6.2.2.6
Visual Effects
To
enter the display visual effects menu, press on “Visual Effects” icon
The Top LED settings allows you to when the Top LED should be illuminated: if you select
Never
, the Top
LED will never illuminate, if you select On Alarm, the Top LED will illuminate only when theH401-T-