5.6 Maintenance Schedule
The unit should be maintained in accordance with the procedures detailed in the
Service Manual. Service maintenance must be performed by a technically competent
individual as described in the Repair Policy.
A. Operator Maintenance
This schedule lists the minimum frequencies. Always follow hospital and local regula-
tions for required frequencies.
Weekly or after each
Clean the Warmer.
patient
After each gas cylinder
Discard old gasket. Use a clean, new gasket.
replacment
Annually
Lubricate the Tee handle screw threads (On models
equipped with gas package).
B. Service Maintenance
This schedule lists the minimum frequencies. Always follow hospital and local regula-
tions for required frequencies.
Per hospital protocol
Perform the electrical safety and checkout procedure from
the Service Manual.
Annually
Check the calibration and perform the service checkout
procedure as detailed in the Service Manual.
Annually
Inspect the yoke strainer nipples and replace if necessary
(On models equipped with gas package).
Every two years
Replace the battery.
Every five years
Replace the regulator(s) (On models equipped with gas
package).
Содержание Ohio 2000 IWS Series
Страница 1: ...Ohio Infant Warmer Systems All Models Operation and Maintenance Manual...
Страница 6: ...iv 6600 0255 000 12 20 00 Table of Contents iv Notes...
Страница 10: ...viii 6600 0255 000 12 20 00 Table of Contents viii Notes...
Страница 12: ...x 6600 0255 000 12 20 00 Table of Contents x Notes...
Страница 16: ...1 4 6600 0255 000 12 20 00 1 Description 1 4 Notes...
Страница 24: ...2 8 6600 0255 000 12 20 00 2 Setup and Checkout Procedures 2 8 Notes...
Страница 76: ...A 18 6600 0255 000 12 20 00 Appendix A 18 Notes...
Страница 77: ...A 19 6600 0255 000 12 20 00 Appendix A 19 Notes...
Страница 78: ...A 20 6600 0255 000 12 20 00 Appendix A 20 Notes...