A-13
6600-0255-000
12/20/00
Appendix
A-13
Table A-1
Nurse Call signals
Warmer
Nurse Call Signal
Status
Pins 1&6
Pins 1&9
Normal
Closed
Open
Alarm
Open
Closed
Power switch off
Open
Closed
or power fails
Nurse Call cable
Open
Open
disconnected
WARNING:
w
If you use the normally open Nurse call connection, a disconnected Nurse
Call cable DOES NOT trigger a remote alarm.
Note: Any interruption of warmer power (deliberately switching off the warmer , acci-
dentally unplugging the power cord, etc.) triggers a Nurse Call alarm.
Installing wall mounted units
WARNING:
w
w
w
w
w
Carefully follow these installation instructions. Failure to do so may
result in serious injury to the operator or patient.
The model 3050, 3100, and 3150 wall mounted warmers were designed for locations where
permanent, infant radiant warmers attached directly to the room wall are desired. To adequately
install these models, pre-planning at the architectural design phase is required.
Drywall
Track
Stud
Sheet Metal
Screws
Ceiling Track
Floor Track
CI.05.040
30 cm
41 cm O.C.
Figure A-1
Wall cross section
Содержание Ohio 2000 IWS Series
Страница 1: ...Ohio Infant Warmer Systems All Models Operation and Maintenance Manual...
Страница 6: ...iv 6600 0255 000 12 20 00 Table of Contents iv Notes...
Страница 10: ...viii 6600 0255 000 12 20 00 Table of Contents viii Notes...
Страница 12: ...x 6600 0255 000 12 20 00 Table of Contents x Notes...
Страница 16: ...1 4 6600 0255 000 12 20 00 1 Description 1 4 Notes...
Страница 24: ...2 8 6600 0255 000 12 20 00 2 Setup and Checkout Procedures 2 8 Notes...
Страница 76: ...A 18 6600 0255 000 12 20 00 Appendix A 18 Notes...
Страница 77: ...A 19 6600 0255 000 12 20 00 Appendix A 19 Notes...
Страница 78: ...A 20 6600 0255 000 12 20 00 Appendix A 20 Notes...