3/Operation
3-7
Error Codes:
In the event of a system failure, a nonsilenceable, high priority alarm sounds and the
System failure LED illuminates. Before shutting off the Warmer and removing it from
service, make note of the error code that appears in the elapsed time display. This code
will aid qualified service personnel in diagnosing the problem. Table 2 lists the possible
error codes, for a more detailed explanation of error codes see the Service manual.
Table 2— Warmer Error Codes
E001
Instruction Test Failure
E002
ADC High Calibration Failure
E003
ADC Low Calibration Failure
E004
Checksum Failure
E005
RAM Test Failure
E007
ADC Converter Failure
E010
Line Voltage Out of Range
E012
Heater Not Switching On
E013
Heater Not Switching Off
E014
Alarm Oscillator Failure
E015
Software Upset - RAM Inconsistency
E016
Safety Relay Not Opening Properly
E017
Software Upset - Missing Routine
E019
Software Upset
E025
Hardware Watchdog Circuit Tripped
3.2 Manual and Servo Mode Operation
WARNING:
w
Before using the Ohmeda Infant Warmer System, read this entire manual.
Attempting to use this device without a thorough understanding of its
operation may result in patient or user injury. This device should only be
operated by personnel trained in its operation and under the direction of
qualified medical personnel familiar with the benefits and risks of this
type of device.
w
Do not leave the patient unattended when using the Infant Warmer. Check
the patient’s temperature periodically using an independent temperature
monitoring device to ensure the comfort and the safety of the patient. If
the warmer is used for an extended time, it is recommended that the
servo mode of operation be used. When an alarm is silenced, close
monitoring of the patient’s condition is required.
w
Use of electrosurgical units or other electrical field radiating equipment
can affect the operation of the Warmer. Keep the patient probe lead as far
away as possible from electrosurgical cables. Do not allow excess electri-
cal cables to be laid on the bed platform. Use of electrosurgical units or
other instruments which radiate electrical fields can cause indirect heat-
ing, by several tenths of a degree of the skin temperature probe due to
absorbed electrical energy. When using these devices near the radiant
warmer, operate the Warmer in manual mode for maximum safety.
w
The use of phototherapy equipment may raise the patient’s temperature.
w
Radiant warmers increase an infant’s insensible water loss. Take appro-
priate measures to maintain the patient’s fluid balance while caring for
them in a radiant warmer.
Содержание Ohio 2000 IWS Series
Страница 1: ...Ohio Infant Warmer Systems All Models Operation and Maintenance Manual...
Страница 6: ...iv 6600 0255 000 12 20 00 Table of Contents iv Notes...
Страница 10: ...viii 6600 0255 000 12 20 00 Table of Contents viii Notes...
Страница 12: ...x 6600 0255 000 12 20 00 Table of Contents x Notes...
Страница 16: ...1 4 6600 0255 000 12 20 00 1 Description 1 4 Notes...
Страница 24: ...2 8 6600 0255 000 12 20 00 2 Setup and Checkout Procedures 2 8 Notes...
Страница 76: ...A 18 6600 0255 000 12 20 00 Appendix A 18 Notes...
Страница 77: ...A 19 6600 0255 000 12 20 00 Appendix A 19 Notes...
Страница 78: ...A 20 6600 0255 000 12 20 00 Appendix A 20 Notes...