2-1
6600-0255-000
12/20/00
2/Setup and Checkout Procedures
2-1
2/Setup and Checkout Procedures
In this section
2.1
Setup ................................................................................................................... 2-1
2.2
Mechanical Checkout Procedure ........................................................................ 2-1
A. Overall Appearance ....................................................................................... 2-1
B. Heater Assembly Rotation ............................................................................. 2-2
C. Mechanical Checks ........................................................................................ 2-2
D. Warmer/Bassinet Interlock (model 3500 only) ............................................... 2-3
E. Warmer/Bassinet Unlock (model 3500 only) ................................................. 2-3
F. Accessory Checks ......................................................................................... 2-3
G. Wall Mount Checkout ..................................................................................... 2-4
2.3
Control Unit Checkout Procedure ....................................................................... 2-4
A. Control Unit Check ......................................................................................... 2-5
B. Elapsed Timer Check .................................................................................... 2-6
C. Observation Light Check ............................................................................... 2-6
D. Raise and Lower Bed Switch Check (Elevating Models Only) ...................... 2-6
E. Interlock Switch Check .................................................................................. 2-7
F. Power Failure, Memory and Battery Test (all models except the
2001 International) ......................................................................................... 2-7
2.1 Setup
Refer to the setup instructions shipped with the warmer for initial unpacking and setup of
the unit after shipment. For model 3050, 3100 and 3150 wall mounted units, refer to the
mounting instructions in the Appendix.
After removal from the shipping containers, inspect the Ohmeda Infant Warmer System and
all accessory items for any signs of damage which may have occurred during shipment. File
a damage claim with the shipping carrier if damage has occurred. Also confirm the pres-
ence of all accessory items or factory installed options as listed on the packing slip.
2.2 Mechanical Checkout Procedure
WARNING:
w
Before using the Ohmeda Infant Warmer System, read this entire manual.
Attempting to use this device without a thorough understanding of its
operation may result in patient or user injury.
w
Do not perform the Checkout Procedures (Mechanical and Control Unit)
while a patient occupies the warmer.
w
Complete the “Checkout Procedures” section of this manual before putting
the unit into operation. If the warmer fails any portion of the checkout
procedures it must be removed from use and repaired.
A. Overall Appearance
1.
Disconnect the power cord from the ac power source for the mechanical checks
portion of this procedure.
2.
Check the overall appearance of the Infant Warmer/ Bassinet System. There should
be no obvious damage.
Содержание Ohio 2000 IWS Series
Страница 1: ...Ohio Infant Warmer Systems All Models Operation and Maintenance Manual...
Страница 6: ...iv 6600 0255 000 12 20 00 Table of Contents iv Notes...
Страница 10: ...viii 6600 0255 000 12 20 00 Table of Contents viii Notes...
Страница 12: ...x 6600 0255 000 12 20 00 Table of Contents x Notes...
Страница 16: ...1 4 6600 0255 000 12 20 00 1 Description 1 4 Notes...
Страница 24: ...2 8 6600 0255 000 12 20 00 2 Setup and Checkout Procedures 2 8 Notes...
Страница 76: ...A 18 6600 0255 000 12 20 00 Appendix A 18 Notes...
Страница 77: ...A 19 6600 0255 000 12 20 00 Appendix A 19 Notes...
Страница 78: ...A 20 6600 0255 000 12 20 00 Appendix A 20 Notes...