A-14
6600-0255-000
12/20/00
Appendix
A-14
C
L
The following instructions describe the manufacturer’s recommended method of installation in
new construction. Use only the Ohmeda hardware provided to mount the model 3050, 3100,
and 3150. The installation should be approved by the appropriate State and Local authorities.
Deviation from these recommendations or attempts to retrofit existing construction should only
be undertaken by professionals experienced in structural design, who in turn are wholly respon-
sible for the structural integrity of the mounting method they devise.
Pre-Installation Preparation
A pattern of two (2) Extra Duty Drywall Studs (STE-20 gauge) and one (1) 10 x 4 cm
Drywall Track (16 gauge) at 41 cm centers form the basic “rough in” for installing each
IWS 3050 unit. See Figure A-1. Verify that the studs and track are securely fastened
together at 30 cm intervals and to both the floor and ceiling tracks with #10 sheet
metal screws.
Warmer Installation
After the wall is finished and the room is completed you are ready to install the warmer.
The distance from the warmer heater head to the bassinet or bed mattress surface is
critical for proper operation of the warmer. This distance must be 69 - 5, +23 cm.
Before installing the model 3050 or 3100, you must determine the bed surface height
that will be used (for your convenience, a bed height label is provided to attach to the
wall to aid in visually aligning the bassinet/bed surface).
CI.05.041
Mounting
Anchors
Hinge
Bracket
Heater
Assembly
88.4 cm
Figure A-2
Warmer mounitng overhead view
1.
Locate the stud edges by pre-measurement or by using a stud finder. Find the center
of the drywall track 86 mm over from the stud edge. See Figure A-2.
2.
Remove the hinge bracket from the warmer by removing the (2) hinge pins (one
on either side) that secure the bracket to the warmer heater head.
41 cm O.C.
102.4 cm
48 cm
Note: For proper ventilation and
access allow at least 10 cm clearance
around the heater assembly.
24 cm
19.8 cm
68.6 cm
All dimensions
are nominal
Содержание Ohio 2000 IWS Series
Страница 1: ...Ohio Infant Warmer Systems All Models Operation and Maintenance Manual...
Страница 6: ...iv 6600 0255 000 12 20 00 Table of Contents iv Notes...
Страница 10: ...viii 6600 0255 000 12 20 00 Table of Contents viii Notes...
Страница 12: ...x 6600 0255 000 12 20 00 Table of Contents x Notes...
Страница 16: ...1 4 6600 0255 000 12 20 00 1 Description 1 4 Notes...
Страница 24: ...2 8 6600 0255 000 12 20 00 2 Setup and Checkout Procedures 2 8 Notes...
Страница 76: ...A 18 6600 0255 000 12 20 00 Appendix A 18 Notes...
Страница 77: ...A 19 6600 0255 000 12 20 00 Appendix A 19 Notes...
Страница 78: ...A 20 6600 0255 000 12 20 00 Appendix A 20 Notes...