1-1
6600-0255-000
12/20/00
1/Description
1-1
1/Description
In this section
1.1
General ............................................................................................................... 1-1
1.2
Support Structure ................................................................................................ 1-1
1.3
Heater Assembly ................................................................................................. 1-1
1.4
Control Unit ......................................................................................................... 1-2
1.5
Bed Platform ....................................................................................................... 1-2
1.6
Bassinet (3500 Series) ........................................................................................ 1-2
1.7
Accessories ......................................................................................................... 1-3
1.1 General
All models of the Ohmeda Medical Infant Warmer System provide a controlled source of
radiant heat for infants and pediatric patients. The control system uses a microprocessor
and provides both manual and servo modes of operation (model 2001 International is
manual mode only). The patient temperature, control temperature, and elapsed time
displays are digital for ease of viewing. The control panel includes a one hour elapsed
timer with time displayed in minutes and seconds. The timer features optional audible
tones for use during Apgar scoring. A complete audio and visual alarm system is in-
cluded on the control panel.
The optional ThermaLink feature simplifies charting and monitoring. RS-232 serial data
is provided for use with customized research software or compatible vital signs monitors.
The Nurse Call connection lets you integrate the Warmer into your current or future
remote alarm systems.
1.2 Support Structure
The support structure provides excellent stability for the radiant heater and optional accessories.
On free standing units, the support structure consists of the base assembly and the uprights with
an integral rail system. The rails provide a means for mounting accessories and ancillary equip-
ment to the unit. Refer to Section 1.7 for a description of the accessory items.
1.3 Heater Assembly
The warmer is designed to warm infants on a bed surface located beneath the heater assembly.
WARNING:
w
Bed-to-heater spacing less than the recommended 64 cm will result in incor-
rect operation and may adversly affect the patient’s condition.
For your convenience, on Model 3000, 3100, 3500, and 4000 the proper bed-to-heater
spacing is indicated by a label located on the right aluminum upright (as viewed from the
front).
Warmers shipped since 1993 have high wattage heaters that allow bed to heater spac-
ing of 69 + 5 - 23 cm, making them compatible with many kinds of bassinets.
Since very low birth weight babies may require tighter control of the heat provided to
them, we recommend that on warmer models that incorporate a built-in bed and are
intended for care of low birth weight babies, the distance between heater and bed be
maintained at 69±5 cm.
Содержание Ohio 2000 IWS Series
Страница 1: ...Ohio Infant Warmer Systems All Models Operation and Maintenance Manual...
Страница 6: ...iv 6600 0255 000 12 20 00 Table of Contents iv Notes...
Страница 10: ...viii 6600 0255 000 12 20 00 Table of Contents viii Notes...
Страница 12: ...x 6600 0255 000 12 20 00 Table of Contents x Notes...
Страница 16: ...1 4 6600 0255 000 12 20 00 1 Description 1 4 Notes...
Страница 24: ...2 8 6600 0255 000 12 20 00 2 Setup and Checkout Procedures 2 8 Notes...
Страница 76: ...A 18 6600 0255 000 12 20 00 Appendix A 18 Notes...
Страница 77: ...A 19 6600 0255 000 12 20 00 Appendix A 19 Notes...
Страница 78: ...A 20 6600 0255 000 12 20 00 Appendix A 20 Notes...