3-1
6600-0255-000
12/20/00
3/Operation
3-1
3/Operation
In this section
3.1
Control Panel Operation ...................................................................................... 3-1
A. Displays ......................................................................................................... 3-1
B. Indicator Lights .............................................................................................. 3-2
C. Switches ........................................................................................................ 3-2
D. Alarms ............................................................................................................ 3-4
3.2
Manual and Servo Mode Operation
................................................................. 3-7
3.3
Elapsed Timer Operation .................................................................................. 3-11
3.4
Bassinet Operation (model 3500 only) ............................................................. 3-11
3.5
Bed Platform Operation ..................................................................................... 3-12
3.6
Side Panel Operation ........................................................................................ 3-13
3.7
X-ray Procedures ..............................................................................................3-14
3.8
Oxygen Administration ...................................................................................... 3-16
3.9
Gas Cylinder Installation and Operation ........................................................... 3-16
3.10 Mounting Accessories .......................................................................................3-18
A. Mounting and Releasing Procedure ............................................................ 3-19
B. Adapter Plate Mounting and Releasing Procedure ...................................... 3-19
C. ECMO Adapter ............................................................................................ 3-20
3.11 ThermaLink Option ............................................................................................ 3-20
3.12 Rotating Drawer Option ..................................................................................... 3-22
3.1 Control Panel Operation
(Control Panel is shown in Figure 2-2)
A. Displays
• Patient Temperature
• Control Temperature
• % Power
• Elapsed Time
Patient Temperature:
The patient temperature display indicates the temperature sensed at the skin tempera-
ture probe tip. The skin temperature probe must be properly connected to the warmer
and the infant for an accurate patient temperature measurement. The range of tem-
perature measurement is from 30.0 to 42.0°C, with a resolution of 0.1°C. Temperatures
above this range result in an “HH.H” patient temperature display and below this range
result in an “LL.L” patient temperature display .
Control Temperature: (all models except the 2001 International)
The control temperature display indicates the control temperature you select in the
servo mode. The servo control temperature range is from 35.0 to 37.5°C. The control
temperature display is not active in the manual mode.
Содержание Ohio 2000 IWS Series
Страница 1: ...Ohio Infant Warmer Systems All Models Operation and Maintenance Manual...
Страница 6: ...iv 6600 0255 000 12 20 00 Table of Contents iv Notes...
Страница 10: ...viii 6600 0255 000 12 20 00 Table of Contents viii Notes...
Страница 12: ...x 6600 0255 000 12 20 00 Table of Contents x Notes...
Страница 16: ...1 4 6600 0255 000 12 20 00 1 Description 1 4 Notes...
Страница 24: ...2 8 6600 0255 000 12 20 00 2 Setup and Checkout Procedures 2 8 Notes...
Страница 76: ...A 18 6600 0255 000 12 20 00 Appendix A 18 Notes...
Страница 77: ...A 19 6600 0255 000 12 20 00 Appendix A 19 Notes...
Страница 78: ...A 20 6600 0255 000 12 20 00 Appendix A 20 Notes...