Refer to the Appendix for additional information on Nurse Call connections.
Nurse Call checkout
1.
Complete the checkout procedure in section 2.3.
2.
Verify proper operation of the Nurse Call station.
3.
Connect the Nurse Call connector to the warmer.
4.
Unplug the warmer to trigger an alarm. Verify that you also get an alarm at the
Nurse Call station.
Note: Any interruption of warmer power (deliberately switching off the warmer, acci-
dently unplugging the power cord, etc.) triggers a Nurse Call alarm.
3.12 Rotating Drawer Option
The rotating drawer option allows access to the storage drawers from the front or from
either side of the warmer. The drawer cabinet rotates by pushing on the side of the
cabinet, stoping at 90° detent positions.
WARNING:
w
Use caution when rotating the cabinet to avoid damage to the drawers or
possible injury. Always ensure the drawers are fully closed before rotat-
ing the cabinet.
Note: The rotating drawer option is not available on model 3400 or 3500 warmers.
Figure 3-13
Rotating Drawer Package
Содержание Ohio 2000 IWS Series
Страница 1: ...Ohio Infant Warmer Systems All Models Operation and Maintenance Manual...
Страница 6: ...iv 6600 0255 000 12 20 00 Table of Contents iv Notes...
Страница 10: ...viii 6600 0255 000 12 20 00 Table of Contents viii Notes...
Страница 12: ...x 6600 0255 000 12 20 00 Table of Contents x Notes...
Страница 16: ...1 4 6600 0255 000 12 20 00 1 Description 1 4 Notes...
Страница 24: ...2 8 6600 0255 000 12 20 00 2 Setup and Checkout Procedures 2 8 Notes...
Страница 76: ...A 18 6600 0255 000 12 20 00 Appendix A 18 Notes...
Страница 77: ...A 19 6600 0255 000 12 20 00 Appendix A 19 Notes...
Страница 78: ...A 20 6600 0255 000 12 20 00 Appendix A 20 Notes...