EN 34
7000 SERIES INDICATORS
When an approval region is selected, the indicator will require that SW1-1 be turned ON before setup can be
exited. If an approval is selected and SW1-1 is not ON, an error message of [Err 0001] will appear when trying to
exit setup.
Certificate #
The Certificate field enables entry of the W&M certificate number for the region. This data is shown on the
Information Recall screen along with other approval information. Enter the certificate number (an alphanumeric
string of up to 16 characters) in the entry box.
Capacity and Increment
Use the Capacity and Increment setup screen to select primary units, set the number of weighing ranges, and the
scale capacity and increment size.
Primary Units
Set the primary units from the selection box choices, which include:
Grams (g)
Pounds (lb)
Tons (ton)
Kilograms (kg)
Tonnes (t)
#Ranges
Set the number of ranges to either one or two in the selection box. If two is selected, the increment size will change
when the weight reaches the second range. The weighing ranges are entered next.
>|1|< Capacity
Specify the weight capacity for range 1. If only one range is enabled, this will be the scale capacity and the display
will show an overcapacity graphic if the weight exceeds this value by more than five increments. If two ranges are
enabled, this will be the weight at which the increment size changes from the range 1 increment size to the range
two increment size.
>|1|< Increment
Specify the increment size for weighing range 1. If only one range is enabled, this will be the increment size for the
entire weighing range of the scale. If two ranges are enabled, this will be the increment used in the lower range.
>|2|< Capacity
Specify the weight capacity for range 2. If only one range is enabled, this parameter will not be shown. If two
ranges are selected, this will be the scale capacity and the display will show an overcapacity graphic if the weight
exceeds this value by more than five increments.
>|2|< Increment
Specify the increment size for weighing range 2. If only one range is enabled, this parameter will not be shown. If
two ranges are selected, this will be the increment size for the second weighing range of the scale.
To use multiple range mode, the ration of range 2 increment to range 1 increment must be less than 50.
Calibration
The Calibration screen enables entry of a GEO code adjustment value, linearity adjustment, and provides access to
zero and span adjustments.
GEO Code
Enter the GEO code for the appropriate adjustment value for the current geographical location. GEO codes are
numbered 0
–31. (Refer to Appendix D.)
www.
GlobalTestSupply
.com
Find Quality Products Online at:
Содержание T72XW AM
Страница 2: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...
Страница 3: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...
Страница 4: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...
Страница 8: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...
Страница 142: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...