EN 92
7000 SERIES INDICATORS
5
SERVICE AND MAINTENANCE
The T72XW indicator is designed to provide years of dependable operation.
However, OHAUS CORPORATION recommends that
– as with any industrial measurement equipment – the
indicator and the connected scale system be serviced periodically.
Timely, factory-specified maintenance and calibration by a OHAUS CORPORATION authorized service technician
will ensure and document accurate and dependable performance to specifications.
5.1 Cleaning and Maintenance
Clean the indicator
’s keypad and cover with a clean, soft cloth that has been dampened with a mild glass cleaner.
Do not use any type of industrial solvent such as toluene or isopropanol (IPA) that could damage the indicator
’s
finish. Do not spray cleaner directly on the indicator.
Regular maintenance inspections and calibration by a qualified service technician are recommended.
The T72XW indicator is a rugged instrument; however, the front panel is a relatively thin covering over sensitive
electronic switches and a lighted display. Care should be taken to avoid any punctures to this surface or any
vibrations or shocks to the instrument. If the front panel is punctured, ensure that steps are taken to prevent dust
and moisture from entering the unit until the indicator can be repaired.
5.2 Service
Only qualified personnel should perform installation, programming, and service. Please contact a local authorized
OHAUS CORPORATION representative for assistance.
OHAUS CORPORATION recommends periodic preventative maintenance to the indicator and scale system to
ensure reliability and to maximize service life. All measurement systems should be periodically calibrated and
certified as required to meet production, industry and regulatory requirements. We can help you maintain uptime,
compliance and quality system documentation with periodic maintenance and calibration services. Contact your
local OHAUS CORPORATION authorized service organization to discuss your requirements.
ONLY PERMIT QUALIFIED PERSONNEL TO SERVICE
THE INDICATOR.EXERCISE CARE WHEN MAKING
CHECKS, TESTS AND ADJUSTMENTS THAT MUST
BE MADE WITH POWER ON.FAILING TO OBSERVE
THESE PRECAUTIONS CAN RESULT IN BODILY
HARM AND/OR PROPERTY DAMAGE.
5.3 Troubleshooting
Troubleshooting activities described here are to assist in identifying whether the problem is in the T72XW indicator
or has an external cause.
Blank Display
On/Off Key Test
Coin Cell Battery Test
Problem Diagnosis
Error Codes and Error Messages
Internal Diagnostic Testing
5.3.1
Blank Display
If the display on the T72XW indicator appears to be blank, first confirm if the display is actually blank or if the
backlight is simply off or if the contrast adjustment is incorrect. Even with the backlight off or a poor contrast setting,
changing weight digits should still be barely visible with a change in weight.
If changing weight digits can still be seen, check the contrast and backlight settings in setup. Moving the indicator
to a very bright environment may help to navigate setup easier.
www.
GlobalTestSupply
.com
Find Quality Products Online at:
Содержание T72XW AM
Страница 2: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...
Страница 3: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...
Страница 4: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...
Страница 8: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...
Страница 142: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...