OHAUS T72XW AM Скачать руководство пользователя страница 102

EN 94 

7000 SERIES INDICATORS 

 

 
Error messages will be shown on the top line of the display and will overwrite the DIO status or time and date 
information if they have been enabled in setup. After the message is cleared, the display reverts to the previous 
data shown before the error was detected. An example of an error message display is given in Figure 5-1 and the 
possible error messages are listed in Table 5-3 with a brief description of the error. 
 

 

 

Figure 5-1: Error Message Display 

 

Table 5-3: Error Codes 

 

Error Message 

Description 

Input string illegal 

An alpha-numeric value was entered where only a 
numeric value is valid 

Preset tare failed 

 

Command failed - . 

 

exceeds capacity limit 

Unit switching would result in a value too large to 
display 

exceeds zero range (+) 

ZERO key pressed and weight is above capture 
range 

exceeds zero range (-) 

ZERO key pressed and weight is below capture 
range 

illegal scale mode 

ZERO key pressed while indicator is in net mode 

illegal tare value 

 

invalid increment value 

 

invalid tare function 

 

invalid zero function 

 

keypad tare disabled 

A preset tare was attempted while the keypad tare 
function is disabled 

motion 

Scale motion inhibited completion of command 

no demand output 

A print was requested without a demand connection 

power-up zero not captured 

 

scale in x10 mode 

Command not possible while scale is expanded x10 

scale not ready 

Print requested and print interlock or auto print 
reset requirements have not been met 

scale over capacity 

Print was requested while scale was over capacity 

scale under zero 

Print was requested while scale was blanked under 
capacity 

tare exceeds limit 

Preset tare value is greater than scale capacity 

tare not enabled 

TARE

 key pressed while tare is disabled 

tare too small 

TARE

 key was pressed while weight was less than 

1 display increment 

tare under zero 

TARE

 key was pressed while weight was below 

zero 

Coin cell voltage low 

Contact service 

 
 

5.3.5 

Internal Diagnostics 

The T72XW indicator provides several internal diagnostic tools that are accessible in setup mode.These tests are 
intended to assist in diagnosing whether a problem is internal or external to the indicator. 
 
To access these tools, enter the Operator menu then the 

SETUP

 mode and navigate to the Maintenance menu. 

Open the Maintenance block then the Diagnostic block. 

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Содержание T72XW AM

Страница 1: ...7000 Series Indicators Instruction Manual T72XW www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Страница 2: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Страница 3: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Страница 4: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Страница 5: ...MENU 16 2 3 1 Language Selection F Codes 16 2 3 2 Menu Navigation 16 2 3 3 Alibi Memory 18 2 3 4 Adjust Contrast 19 2 3 5 Transaction Counter 19 2 3 6 Totals Memory 20 2 3 7 Expand x 10 20 2 3 8 Infor...

Страница 6: ...82 4 5 1 Overview 82 4 5 2 Temporary Tare Weighing 82 4 5 3 Permanent Tare Weighing 83 4 5 4 Operational Features 83 4 5 5 Configuration 83 4 5 6 Operation 84 4 5 7 General 85 4 5 8 Temporary ID Table...

Страница 7: ...DICATORS EN 3 APPENDIX A DEFAULT TEMPLATES 102 APPENDIX B ALIBI MEMORY 104 APPENDIX C COMMUNICATIONS 107 APPENDIX D GEO CODES 134 www GlobalTestSupply com Find Quality Products Online at sales GlobalT...

Страница 8: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Страница 9: ...for the following option boards Choice of one serial DIO option COM2 Serial Interface USB and Discrete I O interface Choice of one network interface Ethernet TCP IP Front panel key access to basic we...

Страница 10: ...s 2 or 3 mV V Number of Scales One Analog Digital Update Rate Internal analog 366 Hz Load Cell Excitation Voltage 10 VDC Minimum Sensitivity 0 1 microvolt per increment Keypad 25 keys polyester overla...

Страница 11: ...n about the product Warning Symbols General Hazard Electrical Shock Hazard 1 5 Safety Precautions Verify that the input voltage range printed on the data label and the plug type matches the local AC p...

Страница 12: ...WAIT AT LEAST THIRTY 30 SECONDS BEFORE ANY CONNECTIONS OR DISCONNECTIONS ARE MADE FAILURE TO OBSERVE THESE PRECAUTIONS COULD RESULT IN DAMAGE TO OR DESTRUCTION OF THE EQUIPMENT AND OR BODILY HARM 1 6...

Страница 13: ...tional and can be configured for various functions such as demand output continuous output extended continuous output SICS host communications or ASCII command input C T P Z The COM2 port is galvanica...

Страница 14: ...tions connect internal wiring and set switches separate the front panel from the enclosure as follows 1 Insert the tip of a flat blade screwdriver into one of the two slots located on the bottom of th...

Страница 15: ...oute the cable through the center of the core and then take one wrap around the outside of the core and route the cable through the center again Either the complete cable or the individual wires can b...

Страница 16: ...erate properly In addition the maximum cable distance must be reviewed Table 1 4 provides recommended maximum cable lengths based on TSR and cable gauge Table 1 4 Recommended Maximum Cable Lengths TSR...

Страница 17: ...e 1 8 Make the connections as necessary NOTES 1 USE ONLY SHIELDED CABLE 2 MAXIMUM CABLE LENGTH 50 FEET 15 METERS 3 WIRE SIZE 18 AWG 823 mm 2 MAXIMUM 24 AWG 0 205 mm2 MINIMUM Figure 1 8 Sample COM1 Con...

Страница 18: ...cator and return all settings back to factory default values This process is described in the T72XW Technical Manual Chapter 4 Service and Maintenance SD Card Installation The SD memory card can be us...

Страница 19: ...nd of each clip engaging Securing the Enclosure When the T72XW indicator is used in a metrologically approved application it must be protected from tampering by use of a seal A wire security seal is i...

Страница 20: ...ne to display the weight units the weight legend the center of zero icon weight range and motion icon Tare values also appear on this line to the left of the display At the bottom of the normal basic...

Страница 21: ...s used for indicating the weight value and other types of information related to the weight Refer to Figure 2 1 Figure 2 1 Elements of the Display The symbols that may appear on the display are descri...

Страница 22: ...ode is available only when entering template strings The current entry mode is indicated at the upper right corner of the display Figure 2 1 Table 2 1 Entering Numeric Data To enter a number into eith...

Страница 23: ...operates regardless of motion on the scale Note that once the tare value has been cleared it cannot be recalled The complete tare process as described above must be performed When in alphanumeric ent...

Страница 24: ...Memory setting time and date adjusting contrast viewing and resetting the transaction counter viewing and clearing totals expanding the displayed weight resolution by 10 and recalling information An...

Страница 25: ...played press ENTER to confirm the selection and move to the next Minutes field label When all the time and date values are correctly set press the LEFT arrow key to return to the Operator Menu Figure...

Страница 26: ...be necessary to enter a valid user name and password to access the setup menus Counting Application Switch Sampling Mode Switches sampling mode between piece sampling and APW entry ID Memory Displays...

Страница 27: ...LIBI SEARCH VIEW screen opens displaying the results Figure 2 5 Alibi Search View Press the UP and DOWN arrow keys to scroll through the entries Press the PRINT key to output the search results to a c...

Страница 28: ...ion prints weights in the expanded format Approved Mode An extra digit of resolution is added to the main weight display in a smaller size e g 123 45 becomes 123 456 The range symbols operate normally...

Страница 29: ...fter a weighing system is installed and is operational it should not be necessary for an operator to access setup Note that a security password can be enabled in setup When a password is set it must b...

Страница 30: ...the weight of an empty container A tare value subtracts from the gross weight measurement providing the computation of the net weight material without the container The tare function can also be used...

Страница 31: ...can be entered in free format If the entered value does not match the displayed weight decimal point location or display interval the entered tare value is rounded to the nearest display interval and...

Страница 32: ...ion is detected after the weight on the scale exceeds a preset tare threshold weight the T72XW will wait for a no motion condition If a stable no motion weight condition occurs within three seconds th...

Страница 33: ...additional digit For example a weight display of 40 96 could increase by one additional digit to display as 40 958 The Expand By 10 mode is indicated on the display by x10 which displays below the wei...

Страница 34: ...displayed in focus in the entry box press the CLEAR key Pressing the ID key last Begin to enter the desired ID value using the alpha numeric keypad A Data prompt will appear at the bottom of the displ...

Страница 35: ...is pressed the indicator will leave normal weighing mode and enter whichever application mode is selected in setup at Application Function Key Assignment Press the FUNCTION key again to return to norm...

Страница 36: ...in setup so that either or both of them is available In this case pressing the FUNCTION key enters the first enabled vehicle weighing mode The next press of the FUNCTION key either returns to normal...

Страница 37: ...ection in Chapter 2 Operation and the Configuration Options Indicator User section of this chapter for further information about password setup and security To return to the OPERATOR menu without ente...

Страница 38: ...y of a parameter value The shortcut method of navigation is not supported on setup pages so traditional navigation must be used Navigate from parameter to parameter by using the DOWN and UP keys then...

Страница 39: ...tering Alphanumeric Data section of Chapter 2 OR Press the LEFT and RIGHT navigation keys to move the cursor into position if the value needs to be edited from a specific point rather than replaced Po...

Страница 40: ...ers in the Scale branch Each of these setup parameters are described in this section If the metrology switch is in the approved position SW1 1 ON access to certain parameters in the Scale branch is re...

Страница 41: ...print template Approval Approval refers to the metrological weights and measures approval configuration for the specific scale The selection list can be set to None no approval is required Canada OIM...

Страница 42: ...value by more than five increments If two ranges are enabled this will be the weight at which the increment size changes from the range 1 increment size to the range two increment size 1 Increment Spe...

Страница 43: ...capture Press the LEFT navigation key if the calibration process must be aborted and the display will return to the previous screen 5 After the first calibration step has completed the menu will eithe...

Страница 44: ...ro Other options are 20d The indicator displays a negative weight until 20 divisions below zero and then blanks the display 20d with Zero The indicator displays a negative weight until 20 divisions be...

Страница 45: ...icator will display and print a negative net weight when the tare weight is larger than the gross weight If net sign correction is Enabled the indicator will switch the gross and tare weight fields on...

Страница 46: ...alue the indicator automatically clears tare and returns to gross mode Motion Check Enable the motion check setting to prevent auto clear from occurring when the scale is in motion below the threshold...

Страница 47: ...elect a values from Disabled 3 10 or 30 seconds the default value being 3 A smaller value means that less time will be used to check for no motion before aborting a command When disabled is selected t...

Страница 48: ...when the scale is in motion below the Reset On return point Motion Check does not appear if Reset on is set to Deviation Scale Reset To initiate a reset of the Scale branch press the ENTER key If the...

Страница 49: ...e use of Alibi Memory requires installation of an SD memory card Attempts to generate a print transaction from the front panel without an SD card installed return a pop up error message SD card is not...

Страница 50: ...f weighments and to be reset while the grand total continues to track the grand total of weight Choose to Clear ST on Print or not by selecting Enabled or Disabled in the selection box Convert Weight...

Страница 51: ...the application setup branches will be shown and the next parameter will be the Reset function refer to page 45 For details on the setup configuration and operation of each of these applications pleas...

Страница 52: ...cation Menu Checkweighing Checkweigh Operation Source Tolerance Type Target Editing Display Mode Motion Blanking Display Quick Set Target Menu Keys Target Values Active Target Target Table Tolerance U...

Страница 53: ...ys If the reset was not successful an error message that reads Reset Failure displays If the reset fails try to initiate the reset again If the reset continues to fail contact a local OHAUS CORPORATIO...

Страница 54: ...g from a to b to c by pressing the key again Once the Key Timing value has elapsed the selected letter is set and the next key press will initiate a new character or digit entry Values between 3 and...

Страница 55: ...are display area Adjust Contrast This parameter accesses a screen where the display contrast can be adjusted Press the UP navigation key to increase the contrast and the DOWN navigation key to decreas...

Страница 56: ...e month Year Use the numeric keypad to enter the year in the Year field text box Language Use the Language setup screen to specify the language for indicator operations Menu Language Use the Menu Lang...

Страница 57: ...of the actual password for access A numeric entry up to six digits in length can be entered Confirm Password Confirm the password previously entered If the password does not agree it will not be acce...

Страница 58: ...the field number from 1 to 25 in the first selection box then select the item for that field in the second selection box Using this method a template of up to 25 fields can be created To terminate a t...

Страница 59: ...e additional items for each template are listed in Chapter 4 Add Line Feeds On each template page is an entry box for entering the number of blank lines to be added at the end of the template This pro...

Страница 60: ...he number of blank lines CR LF to be placed at the start of each report Record Separator A repeated character may be selected as a separator between printed records in the report This step selects the...

Страница 61: ...9 Reports Table 3 6 COM2 Connection Assignments Port Assignment Template Checksum COM2 Assignment Continuous Output Disabled Enabled Continuous Extended Disabled Enabled Demand Template 1 2 3 4 9 Repo...

Страница 62: ...int Client assignment is available only for the Ethernet port If selected as the Ethernet port assignment a Print Client port will be shown to select the connections for the Print Client The Template...

Страница 63: ...ameters for the COM2 and USB ports Baud Use the Baud selection box to set the baud rate for the serial port Options are 300 1200 4800 19200 57600 600 2400 9600 38400 115200 Data Bits Use the Data Bits...

Страница 64: ...proceed to the next group The default value for the gateway is blank After entry is complete press the LEFT navigation key to return to the setup menu Print Client The Print Client connection allows t...

Страница 65: ...oad Cell Output The Load Cell Output screen displays the current number of counts active value for the scale These are raw counts and do not take into consideration the zero or span factors Press the...

Страница 66: ...ce is connected to the receiving port any ASCII data received displays in the receiving field Discrete I O Test Discrete I O Test setup screens include testing for both inputs and outputs VERY IMPORTA...

Страница 67: ...meters in the indicator except metrological significant settings such as Scale type capacity etc When the Reset All screen is first accessed a message displays that asks for verification to reset all...

Страница 68: ...es to factory default settings 6 A status message appears that verifies a successful reset 7 Press the LEFT navigation key to return to the setup menu display 8 This same procedure can be used to rese...

Страница 69: ...riod Display of remaining time in cycle Final average weight display and print When the FUNCTION key is pressed from the basic weigh mode the initial Animal Dynamic weighing display screen will be sho...

Страница 70: ...etup steps for the Animal Dynamic weighing application When the FUNCTION key is properly programmed the Animal Dynamic weighing branch of setup is displayed as shown in Figure 4 1 Figure 4 1 Animal Dy...

Страница 71: ...t is selected from a drop down list The selection list varies depending on the Mode selected Selection Function Mode 1 Mode 2 Disabled Line 2 is blank Average Not available Line 2 shows Average weight...

Страница 72: ...s enabled the cycle will start automatically when the weight on the scale exceeds the programmed threshold and there is no motion on the scale 7 The weight display will show dashes and Line 1 will ind...

Страница 73: ...fields of data supplied by the indicator Figure 4 3Figure 4 3 shows an example of the printed default template String 1 New Line String 2 New Line Time 3 spaces 3 spaces Date New Line ID New Line Aver...

Страница 74: ...is pressed from the basic weigh mode the initial Checkweighing display screen will be shown The icon in the lower left corner indicates the Checkweighing application mode Pressing the FUNCTION key ag...

Страница 75: ...tolerance value The Over Zone output turns on when the weight on the scale exceeds the target plus the tolerance value 4 3 3 Configuration The FUNCTION key must be programmed for Checkweighing in set...

Страница 76: ...isabled none of the values can be edited in the Operator Menu Hold Timer The Hold Timer is a value between 0 0 and 9 9 seconds that sets the amount of time for which the display will freeze after a pr...

Страница 77: ...nu keys will prohibit an operator from viewing or changing the target and tolerance values These values can still be entered and edited in setup at the Target Values page Target Table The Target Table...

Страница 78: ...nclude a count n and Total value ID NEW Screen The following options are available when setting up a new ID and when editing an existing ID The record is updated when the LEFT navigation key is presse...

Страница 79: ...to Print is enabled the indicator will automatically trigger a demand output when it determines a no motion condition 9 Remove the product from the scale 10 If checkweighing the same product again pla...

Страница 80: ...ght Deviation Field Explanation Target Weight value of target Tolerance Lower and upper tolerance limits expressed as a percentage or as a weight Tolerance Description A descriptive label for display...

Страница 81: ...ge by pressing ENTER The display will return to the checkweighing application using the previous target parameters Setup Mode Target Entry This entry mode is described in the Target Values configurati...

Страница 82: ...he wide format report The printed examples show only the format for the Target Deviation tolerance type TARGET TABLE REPORT 12 27 43 26 Jan 2012 L5 5 L5 Stake 5Pk 26 56 lb 0 22 2 36 12 318 42 L7 5 L7...

Страница 83: ...re In the Sample Tare sequence when the Tare prompt is displayed on the indicator enter the preset tare value using the numeric keypad then press ENTER If the indicator is already in counting mode a p...

Страница 84: ...n the two lower lines of the display Both lines have the same options Disabled the default ID Description and APW Menu Keys The Menu Keys parameters determine which functions will be accessible from t...

Страница 85: ...The following functions are accessible from this view Key Function Opens ID NEW screen where a new ID can be set up CLEAR Prompts Delete record Press ENTER to confirm and delete the displayed record...

Страница 86: ...container from the scale and the display will show a piece count equivalent to the tare weight 10 If counting the same pieces again place an empty container on the scale and press TARE to tare to net...

Страница 87: ...full container on the scale and press ENTER 3 The indicator will display a count of zero pieces 4 As pieces are removed from the container the count will update in a negative direction Counting a Ful...

Страница 88: ...f so configured in setup an accumulator and counter An ID Table record can be recalled by one of two methods If the ID is not known the record can be found by viewing the ID Table If the ID is known i...

Страница 89: ...lements in braces are fields of data supplied by the indicator Figure 4 13 shows a printed sample of the default template String 1 New line String 2 New line Time 3 spaces 3 spaces Date New line Displ...

Страница 90: ...e puts the indicator in Temporary Tare weighing mode and the temporary tare icon will be shown in the lower left corner of the display A second press accesses Permanent Tare weighing and the Permanent...

Страница 91: ...to retrieve records from the Permanent ID table or the Temporary ID table Net Sign Correction When enabled in setup at Scale Tare Types Net Sign Correction allows the T72XW to be used for both shippin...

Страница 92: ...of its weighing cycle Operator Clearing Operator Clearing is shown only if Temporary ID is Enabled This setting can be Enabled or Disabled the default When Operator Clearing is Enabled Temporary ID r...

Страница 93: ...ally triggered An inbound or outbound weighing process cannot start until the weight exceeds this value Transaction Table The Transaction Table can be Disabled the default or Enabled If Enabled this t...

Страница 94: ...iable Count and Total fields Items marked with an asterisk appear only if enabled in setup When editing of the record is complete press the LEFT arrow key to exit the edit screen and save the record C...

Страница 95: ...xx The display will change to the net mode If net sign correction is enabled the net weight will be positive even if the inbound weight is larger than the outbound weight If disabled the net weight ca...

Страница 96: ...setup 2 A vehicle with a previously stored tare weight of 00 pulls onto the scale and the weight exceeds the threshold value 3 When a no motion condition is detected the display will prompt ID 4 Enter...

Страница 97: ...h Temporary and Permanent ID modes are enabled to return to basic weighing mode 2 If an ID string is to be included in the transaction press the ID key then use the alphanumeric keypad to enter the st...

Страница 98: ...e Temporary Tare ID Table and the Permanent Tare ID Table can be configured for either narrow 40 characters or wide 80 characters in setup at Communication Reports The examples shown here include a he...

Страница 99: ...PERMANENT ID REPORT 12 34 55 16 Feb 2012 BRM 3885 Robert Ledo Inc Red 567 18460 lb n 7 Total 102640 Copper Pipe JFB 4850 Sam s Hauling Black 12 14800 lb n 32 Total 293820 Copper Wire BJA 5896 APEX Tru...

Страница 100: ...ssistance OHAUS CORPORATION recommends periodic preventative maintenance to the indicator and scale system to ensure reliability and to maximize service life All measurement systems should be periodic...

Страница 101: ...mis wiring Contact service representative No serial port communications Use serial diagnostics section of Maintenance block of setup to determine if the problem is internal or external to the indicato...

Страница 102: ...le the keypad tare function is disabled motion Scale motion inhibited completion of command no demand output A print was requested without a demand connection power up zero not captured scale in x10 m...

Страница 103: ...ale platform the number of counts should increase Press the LEFT navigation key to exit the display Calibration Values Displays the current calibration values for the scale If these values are recorde...

Страница 104: ...Transmit and Receive Indicators 5 To select another port for the serial port test press the ENTER key to move focus back into the port selection box select another port 6 To stop the test routine and...

Страница 105: ...rned off again 5 4 System Backup and Restore The optional SD memory card provides a medium to store a setup file containing the indicator s configuration All setup parameters are saved in a file named...

Страница 106: ...emory card if not already installed Refer to Figure 5 7 for location of SD memory card 2 Apply power enter setup and access the restore procedure at Maintenance Restore from SD 3 At the Dataset Name p...

Страница 107: ...4 must be set to ON to reset EEPROM data If SW1 4 is set to OFF metrological EEPROM data will not be affected by the master reset and the original calibration will persist 4 Apply AC power During the...

Страница 108: ...ctly observe the re verification period and inform the weights and measures authorities As verification requirements vary by jurisdiction the purchaser should contact their local weights and measures...

Страница 109: ...safe removal of the batteries at end of life by a waste treatment facility Please dispose of this product in accordance with local regulations at the collecting point specified for electrical and elec...

Страница 110: ...Line 3 String 2 4 New Line 5 Time 6 3 spaces 7 3 spaces 8 Date 9 New Line 10 Transaction 11 New Line 12 Target Description 13 New Line Template 2 Multiple Line Scale ID T D G T N Element Data 1 Scale...

Страница 111: ...12 Average Weight 13 New Line 14 End of template Counting Application Template 6 Counting Element Data 1 String 1 2 New Line 3 String 2 4 New Line 5 Time 6 3 spaces 7 3 spaces 8 Date 9 New Line 10 Di...

Страница 112: ...in indicator setup to activate the transaction counter value Gross or net weight tare weight type of tare and weight unit Viewing Alibi Memory Records Alibi Memory records can be viewed from the icon...

Страница 113: ...efault find all character the asterisk to view all records 3 With the Alibi Memory search results displayed press the PRINT key 4 A Reports connection must be configured to enable the report print Ali...

Страница 114: ...ple ALIBI MEMORY REPORT 12 27 43 15 Jun 2011 14 Jun 2011 16 07 45 0000027 8 2 lb 0 lb T 8 2 lb 14 Jun 2011 14 09 32 0000026 72 7 lb 0 lb T 72 7 lb 14 Jun 2011 11 14 16 0000025 33 3 kg 11 6 kg T 21 7 k...

Страница 115: ...emand Output Mode The demand output mode transmits data only when the indicator receives a print request Print requests are sent to the T72XW indicator when The operator presses PRINT A discrete input...

Страница 116: ...ud the output rate is only approximately 1 per second The data consists of 17 or 18 bytes as shown in Table C 1 Non significant weight data and tare data digits are transmitted as spaces The continuou...

Страница 117: ...0 0 0 lb or kg selected by Status Byte B bit 4 0 0 1 grams g 0 1 0 metrics tons t 0 1 1 ounces oz 1 0 0 not used 1 0 1 not used 1 1 1 tons ton 1 1 1 no units Bit 3 Print Request 1 Bit 4 Expand Data x...

Страница 118: ...shown in this section a space is represented as _ Each command must be terminated by CR LF 0Dxh 0Axh The characters CR and LF which can be input using the ENTER or RETURN key of most PC indicator key...

Страница 119: ...lows ID Response identification __ Space 20xh Status Status of the indicator See description of the commands and responses Parameters Command dependent response code CR Carriage Return 0Dxh LF Line Fe...

Страница 120: ...ntered with commands or sent with responses The following commands of MT SICS level 0 are supported I0 Inquiry of all implemented MT SICS commands I1 Inquiry of MT SICS level and MT SICS versions I2 I...

Страница 121: ...0_kg T72XW Model number of indicator Standard Standard functionality 50 00 kg Capacity and primary unit of the scale Response I2_I Command understood not executable at present Comments The number of c...

Страница 122: ...overload range Response S_ Indicator in underload range Comments The response to the command SI is the last internal weight value stable or dynamic before receipt of the SI command SIR SEND WEIGHT VA...

Страница 123: ...d The Reset command is always executed except that a Reset command received by the indicator during the calibration and test procedure cannot be processed Commands and Responses MT SICS Level 1 The fo...

Страница 124: ...e exceeded Comments The existing tare will be overwritten and replaced by the new preset tare weight value The indicator will wait for up to 3 seconds after receiving a T command for no motion If moti...

Страница 125: ...neral format These formats are fixed and cannot be changed A header and a footer consisting of additional line feeds and the record separator character are also selected in the Reports sub block of Co...

Страница 126: ...uses in the indicator are available through Backup to SD Memory Card function Setup Scale read and write Index Name Description 101 Scale Name 20 characters max No entry on indicator 102 Approval Type...

Страница 127: ...109 Range 2 increment size 0 19 Refer to values for 107 110 GEO code 0 37 manual entry 111 Linearity 0 Disabled 1 Enabled 112 Auto Zero 0 Disabled 1 Gross 2 Gross Net 113 Auto Zero Range 0 0 5d 1 1d...

Страница 128: ...ton 129 Low Pass Filter 0 Very light 1 Light 2 Medium 3 Heavy 130 Stability Filter 0 Disabled 1 Enabled 131 Motion Range 0 Disabled 1 1d 2 3d 132 No motion interval 0 0 3 s 1 0 5 s 2 0 7 s 3 1 s 133 T...

Страница 129: ...6 SICS SIR command 7 Tare 8 Unit switch 9 Zero 10 Start animal weigh 11 Start peak weigh 213 Input 2 Polarity 0 True 1 True 214 Input 2 Assignment Refer to values for 212 215 Output 1 Assignment 0 No...

Страница 130: ...isabled 1 ID 2 Description 3 APW 233 Counting Application Menu Keys ID Table 0 Disabled 1 Enabled 234 Counting Application Menu Keys Reports 0 Disabled 1 Enabled 235 Counting Application Menu Keys Sam...

Страница 131: ...bled 254 Target Table Menu Key 0 Disabled 1 Enabled 255 261 Not usable 262 Vehicle Operation Temporary ID 0 Disabled 1 Enabled 263 Vehicle Operation Auto ID 0 Disabled 1 Enabled 264 Vehicle Operation...

Страница 132: ...2 30 minutes 3 60 minutes 305 System line 0 Blank 1 Discrete I O 2 Time Date 3 DIO and Time Date 306 Tare display 0 Disabled 1 When active 2 Always 307 Time Format 0 12 MM 1 12 MM SS 2 24 MM 3 24 MM S...

Страница 133: ...digits 324 Operator menu Alibi Memory 0 Disabled 1 Enabled 325 Operator menu Contrast Adjust 0 Disabled 1 Enabled 326 Operator menu Transaction Counter 0 Disabled 1 Enabled 327 Operator menu Time Dat...

Страница 134: ...values for 434 440 COM2 Assignment 0 None 1 OH Continuous Output 2 MT Continuous Output 3 Demand Output 4 Reports 5 SICS 441 COM2 template Refer to values for 434 442 COM2 Checksum Refer to values fo...

Страница 135: ...rint Client Template3 Refer to values for 434 460 USB Assignment 0 None 1 OH Continuous Output 2 MT Continuous Output 3 Demand Output 4 Reports 5 SICS 461 USB template Refer to values for 434 462 USB...

Страница 136: ...unter 502 Scale Overloads Counter 503 Not usable 504 Zero Commands Counter 505 Zero Command Failure Counter Calibration Values read and write Index Name Description 601 Zero Counts Zero counts 602 Mid...

Страница 137: ...Target Description 29 Target ID 30 Target Weight 31 Target Tolerances 32 Total 33 Zone 34 APW 35 PIECE_COUNT 36 Record ID 37 Record Description 38 Transaction Date 39 Transaction Time 40 Transaction T...

Страница 138: ...28 Output Template 8 Field Content Array Refer to values for 712 729 Output Template 9 Add Line Feeds Manual Entry 730 Output Template 9 Field Content Array Refer to values for 712 741 Template String...

Страница 139: ...le D 1 on the following pages to determine the GEO code for the new altitude and location at which the scale will be used 2 Enter that GEO value into the GEO code parameter in Setup at Scale Calibrati...

Страница 140: ...8 18 17 17 16 16 15 15 14 14 13 46 45 47 51 19 18 18 17 17 16 16 15 15 14 14 47 51 48 58 19 19 18 18 17 17 16 16 15 15 14 48 58 50 6 20 19 19 18 18 17 17 16 16 15 15 50 6 51 13 20 20 19 19 18 18 17 17...

Страница 141: ...osed to radioactive or corrosive materials has foreign material penetrating to the inside of the product or as a result of service or modification by other than OHAUS In lieu of a properly returned wa...

Страница 142: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Страница 143: ...or trademarks of Microsoft Corporation in the United States and or other countries OHAUS Corporation 7 Campus Drive Suite 310 Parsippany NJ 07054 USA Tel 1 973 377 9000 Fax 1 973 944 7177 With office...

Отзывы: