![OHAUS T72XW AM Скачать руководство пользователя страница 124](http://html1.mh-extra.com/html/ohaus/t72xw-am/t72xw-am_instruction-manual_3537195124.webp)
EN 116
7000 SERIES INDICATORS
Comments
Command to send the current stable weight once then continuously after every weight change greater or equal to
the “value” a non-stable (dynamic) value followed by the next stable value. If no preset value is included, the
weight change must be at least 12.5% of the last stable weight value an a minimum of 30d.
SR is overwritten and cancelled by the commands S, SI, SIR, @ and hardware break.
If, following a non-stable (dynamic) weight value, stability has not been reached within the 3 second timeout
interva
l, the response “S_I “ is sent and then a non-stable weight value. Timeout then starts again form the
beginning.
The range value must be entered in primary units and must be in the range of 1d to capacity.
T
– TARE
Command:
T
Response:
T_S_ _ _ _100.00_kg
Tare performed, meaning the scale was stable and weight was within the
weighing range.
Response:
T_I
Command understood, not executable at present.
Response:
T_+
Upper limit of zero setting range exceeded.
Response:
T_-
Lower limit of zero setting range exceeded.
Comments
The existing tare will be overwritten and replaced by the new preset tare weight value.
The indicator
will wait for up to 3 seconds after receiving a “T” command for no-motion. If motion does not settle
within this time, the command is aborted and the T_I response is sent.
TA
– INQUIRE/ENTER TARE VALUE
Command:
TA
Inquiry of tare weight value
TA_Tare Preset Value_Unit
Entry of a preset tare value.
Response:
TA_A_TareWeightValue_Unit
Current Tare weight value.
Response:
TA_I
Current Tare weight value cannot be transferred (the indicator is currently executing
another command, such as zero setting).
Response:
TA_L
Command understood, invalid parameter.
Example
Command:
TA_10.00_kg
Load a preset tare of 10 kg.
Response:
TA_A_ _ _ _ _10.00_kg
The 10.00 kg tare value was accepted.
Comments
The existing tare will be overwritten and replaced by the new preset tare weight value.
The indicator will automatically round the tare value to the current readability.
The preset value must be entered in primary units.
TAC
– CLEAR TARE VALUE
Command:
TAC
Response:
TAC_A Tare value cleared.
Response:
TAC_I Command understood, not executable at present.
www.
GlobalTestSupply
.com
Find Quality Products Online at:
Содержание T72XW AM
Страница 2: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...
Страница 3: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...
Страница 4: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...
Страница 8: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...
Страница 142: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...