![OHAUS T72XW AM Скачать руководство пользователя страница 123](http://html1.mh-extra.com/html/ohaus/t72xw-am/t72xw-am_instruction-manual_3537195123.webp)
7000 SERIES INDICATORS
EN 115
Z
– ZERO
Command:
Z
Response:
Z_A
Command performed, meaning the scale was in gross mode, scale was stable and weight
was within the zero capture range.
Response:
Z_I
Command understood, not executable at present.
Response:
Z_+
Upper limit of zero setting range exceeded.
Response:
Z_-
Lower limit of zero setting range exceeded.
Comments
The calibrated zero point determined during calibration is not influenced by this command.
The indicator
will wait for up to 3 seconds after receiving an “Z” command for no-motion. If motion does not settle
within this time, the command is aborted and the Z_I response is sent.
@
– RESET
Command:
@
Response:
I4_A_“12345678”
12345678
Serial number of the scale
Comments
Resets the scale to the condition found after switching on, but without a zero setting being performed.
All commands awaiting responses are canceled.
The tare register is cleared.
The SIR and SR commands are cancelled.
The Reset command is always executed except that a Reset command received by the indicator during the
calibration and test procedure cannot be processed.
Commands and Responses MT-SICS Level 1
The following commands of MT-SICS level 1 are available:
SR
Send weight value on weight change (Send and Repeat)
T
Tare
TA
Set or inquiry of preset tare value
TAC
Clear tare value
TI
Tare Immediately
SR
– SEND WEIGHT VALUE ON WEIGHT CHANGE (SEND AND REPEAT)
Command:
SR_Value_Unit
SR
Response:
S_S_ _ _ _ _ 105.1_lb Current stable weight.
S_D_ _ _ _ _ 106.7_lb Dynamic weight value.
S_S_ _ _ _ _ 124.3_lb Next stable weight value.
Response:
S_I
Command understood, not executable at present.
Response:
S_L
Command understood, invalid parameter.
Response:
S_+
Indicator in overload range.
Response:
S_-
Indicator in underload range.
Example
Command:
SR_0.50_kg
Send the current stable weight value followed by every load change > 0.50 kg.
Response:
S_S_ _ _ _ _100.00_kg
Scale stable.
S_D_ _ _ _ _115.23_kg
More than 0.50 kg loaded.
S_S_ _ _ _ _200.00_kg
Scale stable again.
www.
GlobalTestSupply
.com
Find Quality Products Online at:
Содержание T72XW AM
Страница 2: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...
Страница 3: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...
Страница 4: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...
Страница 8: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...
Страница 142: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...