3.3 Cautions on Floating Ground Connection
WF1967/WF1968
3-18
3.3
Cautions on Floating Ground Connection
The signal ground of the BNC terminals for FCTN OUT, SYNC/SUB OUT, and MOD/ADD IN is
shared, but since it is insulated from the enclosure (ground potential), it can be connected to the
equipment that has a different potential. Moreover, when the equipment is mounted in a rack, the signal
ground is not affected by the potential of the rack.
In WF1968, the above-mentioned BNC terminals are also insulated between channels.
Further, the signal ground of the 10MHz REN IN is also insulated from the enclosure. Therefore, noise
caused by ground-loop does not affect the connection with a frequency standard.
Nor does noise caused by ground-loop affect the connection even when synchronously connecting
multiple units of WF1967 and WF1968.
Note that, in all cases, the floating voltage should be limited to 42Vpk (DC + AC peak) or lower to
prevent electric shocks.
The other signal grounds are all connected to the enclosure. The enclosure itself is connected to a
protective conductor terminal of the power supply input.
To prevent electric shocks, do not apply a voltage exceeding 42Vpk
(DC + AC peak) between the ground of the BNC connectors insulated
from the enclosure and the enclosure.
Also, do not apply a voltage exceeding 42Vpk (DC + AC peak)
between the grounds of the BNC connector groups insulated from the
enclosure. "BNC connector groups" indicate multiple BNC connectors
that are connected to a common ground.
If such a high voltage is applied, the internal voltage limiting elements
will try to reduce the voltage, but a too high voltage may cause the
product to be burned.
If a difference in potential exists between the ground of a BNC connector
insulated from the enclosure and the enclosure, do not short-circuit the
hot side of that BNC connector and the enclosure. This may damage the
product.
If a difference in potential exists between the grounds of BNC
connectors, do not short circuit these BNC connector grounds. This may
damage the product.
CAUTION
CAUTION
WARNING
Содержание WAVE FACTORY WF1967
Страница 1: ...MULTIFUNCTION GENERATOR WF1967 WF1968 Instruction Manual Operations NF Corporation nbn Austria GmbH...
Страница 2: ......
Страница 3: ...MULTIFUNCTION GENERATOR WF1967 WF1968 Instruction Manual Operations DA00048287 002...
Страница 4: ......
Страница 10: ...WF1967 WF1968 vi...
Страница 19: ...WF1967 WF1968 1 1 1 Overview 1 1 Features 1 2 1 2 Operating Principles 1 4...
Страница 50: ...3 3 Cautions on Floating Ground Connection WF1967 WF1968 3 20 MEMO...
Страница 156: ...4 9 Setting and Operation of Burst WF1967 WF1968 4 106 MEMO...
Страница 206: ...7 9 Identifying Memory Space Required for Saving Arbitrary Waveforms WF1967 WF1968 7 14 MEMO...
Страница 221: ...WF1967 WF1968 9 1 9 Synchronizing Multiple Units 9 1 Connection Procedure 9 2 9 2 Performing Synchronization 9 4...
Страница 226: ...9 2 Performing Synchronization WF1967 WF1968 9 6 MEMO...
Страница 262: ...13 6 Product Information Display Information WF1967 WF1968 13 4 MEMO...
Страница 274: ...14 7 Suspected Failure WF1967 WF1968 14 12 MEMO...
Страница 275: ...WF1967 WF1968 15 1 15 Maintenance 15 1 Outline 15 2 15 2 Operation Inspection 15 4 15 3 Performance Test 15 5...
Страница 286: ...15 3 Performance Test WF1967 WF1968 15 12 MEMO...
Страница 287: ...WF1967 WF1968 16 1 16 List of Initial Settings...
Страница 326: ......
Страница 328: ......
Страница 329: ......