15.3 Performance Test
WF1967/WF1968
15-6
15.3.2 Sine Wave Amplitude Accuracy Test
Connection:
FCTN OUT → Digital multi-meter (AC voltage TrueRMS measurement)
Use a coaxial cable.
Setting:
The following table shows the amplitude after setting initialization. (Frequency
is set to 1kHz).
Measurement: Measure the output voltage for each waveform as the effective value.
Judgment:
It is normal when the value falls within the following table.
Amplitude setting (load open value)
Rating range
20Vp-p (7.071Vrms)
7.000Vrms to 7.142Vrms
5Vp-p (1.768Vrms)
1.749Vrms to 1.786Vrms
3Vp-p (1.061Vrms)
1.049Vrms to 1.072Vrms
1Vp-p (353.6mVrms)
349.3Vrms to 357.8mVrms
0.3Vp-p (106.1mVrms)
104.3mVrms to 107.8mVrms
0.1Vp-p (35.36mVrms)
34.29mVrms to 36.42mVrms
0.02Vp-p (7.071mVrms)
6.293mVrms to 7.849mVrms
15.3.3 DC Offset Accuracy Test
DC only
Connection:
FCTN
OUT → Digital multi-meter (DC voltage measurement)
Setting:
Set the waveform to DC after setting initialization. The following table shows
the DC offset.
Measurement: Measure the output voltage as DC.
Judgment:
It is normal when the value falls within the following table.
DC offset setting (load open
value)
Rating range
±10V
±9.890V to ±10.110V
±3V
±2.960V to ±3.040V
±1V
±0.9800V to ±1.0200V
±0.3V
±0.2870V to ±0.3130V
0V
-10.00mV to +10.00mV
AC+DC
Connection:
FCTN OUT → Digital multi-meter (DC voltage measurement)
Setting:
Set the oscillation mode to BURST, burst mode to GATE, trigger to external
Off (the oscillation of sine wave is stopped at 0 degree) after setting
initialization.
The following table shows the amplitude. DC offset setting is left to 0 V.
Measurement: Measure the output voltage as DC.
Judgment:
It is normal when the value falls within the following table.
Amplitude setting (load open value)
Rating range
6.4Vp-p
-42.00mV to +42.00mV
3.5Vp-p
-27.50mV to +27.50mV
0.7Vp-p
-13.50mV to +13.50mV
Содержание WAVE FACTORY WF1967
Страница 1: ...MULTIFUNCTION GENERATOR WF1967 WF1968 Instruction Manual Operations NF Corporation nbn Austria GmbH...
Страница 2: ......
Страница 3: ...MULTIFUNCTION GENERATOR WF1967 WF1968 Instruction Manual Operations DA00048287 002...
Страница 4: ......
Страница 10: ...WF1967 WF1968 vi...
Страница 19: ...WF1967 WF1968 1 1 1 Overview 1 1 Features 1 2 1 2 Operating Principles 1 4...
Страница 50: ...3 3 Cautions on Floating Ground Connection WF1967 WF1968 3 20 MEMO...
Страница 156: ...4 9 Setting and Operation of Burst WF1967 WF1968 4 106 MEMO...
Страница 206: ...7 9 Identifying Memory Space Required for Saving Arbitrary Waveforms WF1967 WF1968 7 14 MEMO...
Страница 221: ...WF1967 WF1968 9 1 9 Synchronizing Multiple Units 9 1 Connection Procedure 9 2 9 2 Performing Synchronization 9 4...
Страница 226: ...9 2 Performing Synchronization WF1967 WF1968 9 6 MEMO...
Страница 262: ...13 6 Product Information Display Information WF1967 WF1968 13 4 MEMO...
Страница 274: ...14 7 Suspected Failure WF1967 WF1968 14 12 MEMO...
Страница 275: ...WF1967 WF1968 15 1 15 Maintenance 15 1 Outline 15 2 15 2 Operation Inspection 15 4 15 3 Performance Test 15 5...
Страница 286: ...15 3 Performance Test WF1967 WF1968 15 12 MEMO...
Страница 287: ...WF1967 WF1968 16 1 16 List of Initial Settings...
Страница 326: ......
Страница 328: ......
Страница 329: ......