15.2 Operation Inspection
WF1967/WF1968
15-4
15.2 Operation Inspection
Check Before Operation inspection
Check followings before the operation inspection.
The power supply voltage is within the rated range.
Ambient temperature is within the range of 0 to +40°C.
Ambient relative humidity is within the range of 5 to 85%RH (furthermore, the absolute
humidity is within the range of 1 to 25g/m
3
).
Non-condensing.
Function check
Check at power-on
Make sure that no error message appears at power-on.
When an error message appears,
P. 14-2.
When the display shows any problems at the power-on, turn the power off, wait for at least 5
seconds and turn the power-on again.
Self-diagnosis
Perform Self Check on the Utility screen.
Check of important functions
To avoid miss-setting, perform initialization first (Perform [Reset] in the Utility screen).
Then, connect the FCTN OUT to the oscilloscope using characteristic impedance
50Ω series coaxial
cable and observe the output.
Change following settings several times and check they work properly. For items that set numeric
values like the frequency, check become more reliable by operating both a numeric keypad and
modify knobs.
• Waveform (shortcut key: FCTN key
FCTN
)
• Frequency (shortcut key: FREQ key
FREQ
)
• Amplitude (shortcut key: AMPTD key
AMPTD
)
• DC offset (shortcut key: OFFSET key
OFFSET
)
• Output on/off (
CH 1
,
CH 2
, or
OUTPUT
)
Check of GPIB, USB and LAN(Option)
Perform some of setting changes performed in the check of important function section on the GPIB,
USB and LAN. Verify that Changes of the output is the same.
In this case, GPIB, USB or LAN is displayed on the status display area on the top of the screen.
Check also that pressing the [Local] soft key while GPIB, USB or LAN is displayed on the status
display area clears the previous status display and returns to the local status (except for during local
lockout condition).
Содержание WAVE FACTORY WF1967
Страница 1: ...MULTIFUNCTION GENERATOR WF1967 WF1968 Instruction Manual Operations NF Corporation nbn Austria GmbH...
Страница 2: ......
Страница 3: ...MULTIFUNCTION GENERATOR WF1967 WF1968 Instruction Manual Operations DA00048287 002...
Страница 4: ......
Страница 10: ...WF1967 WF1968 vi...
Страница 19: ...WF1967 WF1968 1 1 1 Overview 1 1 Features 1 2 1 2 Operating Principles 1 4...
Страница 50: ...3 3 Cautions on Floating Ground Connection WF1967 WF1968 3 20 MEMO...
Страница 156: ...4 9 Setting and Operation of Burst WF1967 WF1968 4 106 MEMO...
Страница 206: ...7 9 Identifying Memory Space Required for Saving Arbitrary Waveforms WF1967 WF1968 7 14 MEMO...
Страница 221: ...WF1967 WF1968 9 1 9 Synchronizing Multiple Units 9 1 Connection Procedure 9 2 9 2 Performing Synchronization 9 4...
Страница 226: ...9 2 Performing Synchronization WF1967 WF1968 9 6 MEMO...
Страница 262: ...13 6 Product Information Display Information WF1967 WF1968 13 4 MEMO...
Страница 274: ...14 7 Suspected Failure WF1967 WF1968 14 12 MEMO...
Страница 275: ...WF1967 WF1968 15 1 15 Maintenance 15 1 Outline 15 2 15 2 Operation Inspection 15 4 15 3 Performance Test 15 5...
Страница 286: ...15 3 Performance Test WF1967 WF1968 15 12 MEMO...
Страница 287: ...WF1967 WF1968 16 1 16 List of Initial Settings...
Страница 326: ......
Страница 328: ......
Страница 329: ......