14.2 Error Messages during Operation
WF1967/WF1968
14-4
22029: Edge time changed due to Width
A given pulse width time cannot be achieved with the current edge time (LE, TE). The edge time
becomes shorter.
The pulse width time takes precedence over the edge time.
22030: Edge time changed due to Duty
A given pulse width duty cannot be achieved with the current edge time (LE, TE). The edge time
becomes shorter.
The pulse width duty takes precedence over the edge time.
22031: Edge time and/or Width changed due to Frequency
The specified frequency cannot be achieved with the current edge time (LE, TE) and pulse width
time. The edge time and the pulse width time are changed.
The Frequency takes precedence over the pulse width time and the edge time.
22032: Edge time and/or Duty changed due to Frequency
The specified frequency cannot be achieved with the current edge time (LE, TE) and pulse width
duty. The edge time and the pulse width duty are changed.
The Frequency takes precedence over the pulse width duty and the edge time.
22034: Selected ARB is missing; Edit Memory ARB assigned
The specified arbitrary wave does not exist. An arbitrary wave in the edit memory is assigned
instead of it.
This error happens when an arbitrary wave used before has been deleted during not used. An
arbitrary wave is identified only with the number.
22035: Too narrow or too wide Duty specified; Amplitude may decrease or pulse may be lost
The duty setting is too narrow or too wide. Amplitude may decrease or pulse may be lost.
This event happens when the pulse width of a square wave with extended duty variable range is
narrower than 8.4ns on the low or high side.
22036
: Frequency reduced due to Function
The frequency is changed to the lower value based on the waveform.
The upper limit of frequency depends on the waveform.
22038
: Duty changed due to Extend-Off
The duty variable range of a square wave is changed to standard. The duty is changed to the value
restricted by the frequency.
When the duty variable range is standard, it is changes based on the frequency.
22039: Duty changed due to Frequency
The duty variable range of a square wave is standard. The duty was changed to the value restricted
by the frequency.
When the duty variable range is standard, it changes based on the frequency.
22040: High level changed due to Low level
The change to the low level also changes the high level based on the possible output range.
22041: Low level changed due to High level
The change to the high level also changes the low level based on the possible output range.
22042: Not acceptable due to another CH limitation
The specified setting cannot be set due to another channel restriction in 2-channel equivalence setting.
22043: Frequency and/or DeltaFreq changed due to Function
The frequency and the frequency difference are changed based on the upper frequency of waveform
when the channel mode is DIFF.
22044: Frequency of CH1 changed due to DeltaFreq
The frequency of CH1 is changed to maintain the frequency difference when the channel mode is
DIFF.
The frequency difference takes precedence over the frequency of CH1.
Содержание WAVE FACTORY WF1967
Страница 1: ...MULTIFUNCTION GENERATOR WF1967 WF1968 Instruction Manual Operations NF Corporation nbn Austria GmbH...
Страница 2: ......
Страница 3: ...MULTIFUNCTION GENERATOR WF1967 WF1968 Instruction Manual Operations DA00048287 002...
Страница 4: ......
Страница 10: ...WF1967 WF1968 vi...
Страница 19: ...WF1967 WF1968 1 1 1 Overview 1 1 Features 1 2 1 2 Operating Principles 1 4...
Страница 50: ...3 3 Cautions on Floating Ground Connection WF1967 WF1968 3 20 MEMO...
Страница 156: ...4 9 Setting and Operation of Burst WF1967 WF1968 4 106 MEMO...
Страница 206: ...7 9 Identifying Memory Space Required for Saving Arbitrary Waveforms WF1967 WF1968 7 14 MEMO...
Страница 221: ...WF1967 WF1968 9 1 9 Synchronizing Multiple Units 9 1 Connection Procedure 9 2 9 2 Performing Synchronization 9 4...
Страница 226: ...9 2 Performing Synchronization WF1967 WF1968 9 6 MEMO...
Страница 262: ...13 6 Product Information Display Information WF1967 WF1968 13 4 MEMO...
Страница 274: ...14 7 Suspected Failure WF1967 WF1968 14 12 MEMO...
Страница 275: ...WF1967 WF1968 15 1 15 Maintenance 15 1 Outline 15 2 15 2 Operation Inspection 15 4 15 3 Performance Test 15 5...
Страница 286: ...15 3 Performance Test WF1967 WF1968 15 12 MEMO...
Страница 287: ...WF1967 WF1968 16 1 16 List of Initial Settings...
Страница 326: ......
Страница 328: ......
Страница 329: ......