44
4.
Note that when the temperature catheter is first connected and in open air (room temperature), the Camino
ICP Monitor will:
•
Sound an alarm; and
•
Display "
Temperature out of accuracy range"
on the status bar.
To silence the alarm temporarily (3 minutes) while the catheter is still in air, press the yellow alarm button.
5.
Verify that the ICP value reads 0 mmHg on the monitor. To verify, press the
Main
tab on the touch screen and
view the ICP value. If the ICP value does not read 0 mmHg:
A. Press the
Scale
button on the
Main
tab to select a waveform range of -10 to 20 mmHg. Note that
pressing the
Scale
button each time advances to the next range.
B. Remove the black zero adjustment tool from the catheter kit.
C. Use the tool to adjust the screw located on the bottom of the catheter’s transducer to increase/decrease
the ICP value.
1
Transducer from catheter
2
Zero adjustment tool
D. Set the ICP value to 0 mmHg.
Содержание Camino ICP
Страница 1: ...1 Camino ICP Monitor Intracranial Pressure and Temperature Monitor User s Manual 025870 Rev 00 01 2019...
Страница 2: ...2...
Страница 12: ...12 Blank page...
Страница 16: ...16 Blank page...
Страница 26: ...26 About the Left Panel Theleftpanel doesnotcontain anyusableconnector ports or buttons...
Страница 56: ...56 Blank page...
Страница 74: ...74 Blank page...
Страница 92: ...92 Blank page...
Страница 100: ...100 Blank page...
Страница 110: ...110 Blank page...
Страница 142: ...142 Blank page...
Страница 146: ...146 Blank page...