105
Problem
Cause
Recommendation
Touch screen freezes
System hardware error
•
Press and hold the power button for several seconds to
perform a forced shutdown of the system.
•
Press the power button to turn on the system.
Camino ICP Monitor does not
power on or off when power
button is pressed.
Firmware error
•
Disconnect AC adapter cord from Camino ICP Monitor and
remove battery.
•
Re-insert battery.
•
Reconnect AC adapter cord to Camino ICP Monitor.
•
Press the power button.
The Mean ICP value on the touch
screen displays two dashes (- -).
Poor connection between Fiber Optic Catheter
Cable and Camino ICP Monitor.
Make sure the cable is properly connected to the Camino ICP
Monitor.
Poor connection between Fiber Optic Catheter
Cable or Flex Extension Cable and the
catheter.
Replace the catheter (see directions for use supplied with
catheter).
Faulty catheter
Replace the catheter (see directions for use supplied with
catheter).
Damaged Fiber Optic Catheter Cable
Replace Fiber Optic Catheter Cable.
Damaged Camino Flex Catheter Cable.
Replace Camino Flex Catheter Cable.
The temperature value on the
touch screen displays two
dashes (- -).
Faulty catheter
Replace the catheter (see directions for use supplied with
catheter).
ICP only catheter in use
Replace the catheter with ICP/temperature catheter (see
directions for use supplied with catheter).
Damaged Fiber Optic Catheter Cable
Replace Fiber Optic Catheter Cable.
Содержание Camino ICP
Страница 1: ...1 Camino ICP Monitor Intracranial Pressure and Temperature Monitor User s Manual 025870 Rev 00 01 2019...
Страница 2: ...2...
Страница 12: ...12 Blank page...
Страница 16: ...16 Blank page...
Страница 26: ...26 About the Left Panel Theleftpanel doesnotcontain anyusableconnector ports or buttons...
Страница 56: ...56 Blank page...
Страница 74: ...74 Blank page...
Страница 92: ...92 Blank page...
Страница 100: ...100 Blank page...
Страница 110: ...110 Blank page...
Страница 142: ...142 Blank page...
Страница 146: ...146 Blank page...