107
Problem
Cause
Recommendation
Discrepancy of temperature
measurements between the
Camino ICP Monitor and the
patient bedside monitor.
Incorrect patient bedside monitor adapter
cable being used.
Verify correct patient bedside monitor cable is being used.
PMIO cable or patient bedside monitor
adapter cable is faulty.
Replace PMIO cable or patient bedside monitor adapter cable.
PMIO cable or patient bedside monitor
adapter cable is loose.
Check cable connections between Camino ICP Monitor and
patient bedside monitor.
Patient bedside monitor has malfunctioned. Refer to manufacturer’s troubleshooting guide for the patient
bedside monitor.
No transfer of trend data via
digital streaming with USB-to-
RS232 adapter cable.
Loose USB-to-RS232 adapter cable.
•
Reconnect USB-to-RS232 adapter cable.
•
On touch screen, press
Trend
tab and verify
Data
Streaming Enabled
message appears.
Faulty USB-to-RS232 adapter cable.
•
Replace USB-to-RS232 adapter cable.
•
On touch screen, press
Trend
tab and verify
Data
Streaming Enabled
message appears.
External monitoring device is not configured
correctly for digital streaming.
See digital streaming requirements in Chapter 6.
Содержание Camino ICP
Страница 1: ...1 Camino ICP Monitor Intracranial Pressure and Temperature Monitor User s Manual 025870 Rev 00 01 2019...
Страница 2: ...2...
Страница 12: ...12 Blank page...
Страница 16: ...16 Blank page...
Страница 26: ...26 About the Left Panel Theleftpanel doesnotcontain anyusableconnector ports or buttons...
Страница 56: ...56 Blank page...
Страница 74: ...74 Blank page...
Страница 92: ...92 Blank page...
Страница 100: ...100 Blank page...
Страница 110: ...110 Blank page...
Страница 142: ...142 Blank page...
Страница 146: ...146 Blank page...