65
Item
Description
4
Restore Defaults
Restores customized settings for the high ICP alarm limit to the factory default settings (20
mmHg). Pressing this button also activates the
Alarm
On
button.
5
ICP
Displays the Mean ICP in mmHg as numerical values.
6
Temperature
Displays tissue temperature in degrees Celsius (°C).
About the High ICP Alarm Limit Calculation
The Camino ICP Monitor bases the high alarm limit on the patient’s Mean ICP value, which is represented by the
numerical ICP value that appears on the touch screen. The high ICP alarm is triggered if the patient’s displayed
Mean ICP value continuously remains above the high ICP alarm limit for a period of 5 seconds.