20
•
To prevent electrical shock, only use the AC power adapter supplied by Natus Neuro (REF # 025430). Using a
different AC power adapter may not provide protection against electric shock.
•
To reduce the risk of electric shock, only connect to IEC 60601 certified bedside monitors.
•
Danger
- Possible explosion hazard if used in the presence of flammable anesthetics. Not suitable for use in the
presence of flammable anesthetics mixture with air, oxygen, or nitrous oxide.
•
Only use Natus Neuro supplied accessories on the Camino ICP Monitor. This applies in particular to catheters,
catheter cables, battery, AC power adapter, and USB-to-R232 adapter cable.
•
If the Camino ICP Monitor loses power and shuts down while it is connected to a patient bedside monitor, do not
use the ICP values on the patient bedside monitor for patient measurements; the ICP values on the patient
bedside monitor will be invalid.
•
Connect the monitor to an AC power supply immediately if the low battery alarm is activated.
•
When using the battery:
- Do not heat above 80 °C.
- Do not open battery.
- Do not dispose of in fire.
- Do not short circuit as battery may ignite, explode, leak, or get hot causing personal injury.
- Replace battery with same part number only (REF # 026950).
- Use of another battery may present a risk of fire or explosion.
•
To prevent injury to the patient, user or other persons, make sure that the battery cover is closed securely during
monitor use.
•
Do not connect a Camino Fiber Optic Catheter and an Camino Flex Catheter simultaneously to the monitor.
The Camino ICP Monitor is designed to report ICP measurements with either the Flex Catheter or Fiber Optic
Catheter connected to the monitor, but not both together.
•
To prevent possible patient injury resulting from incorrect ICP measurements, always perform the steps in
Chapter 3 before implanting a new Camino Fiber Optic Catheter into the patient. In particular, always make
sure to adjust the new Fiber Optic Catheter’s ICP value to zero before implantation while the catheter is in the
air. Never attempt to re-zero a catheter while the catheter is inside the patient.
Содержание Camino ICP
Страница 1: ...1 Camino ICP Monitor Intracranial Pressure and Temperature Monitor User s Manual 025870 Rev 00 01 2019...
Страница 2: ...2...
Страница 12: ...12 Blank page...
Страница 16: ...16 Blank page...
Страница 26: ...26 About the Left Panel Theleftpanel doesnotcontain anyusableconnector ports or buttons...
Страница 56: ...56 Blank page...
Страница 74: ...74 Blank page...
Страница 92: ...92 Blank page...
Страница 100: ...100 Blank page...
Страница 110: ...110 Blank page...
Страница 142: ...142 Blank page...
Страница 146: ...146 Blank page...