104
Responding to Problems During Use
The following table provides guidelines for responding to technical problems that may occur with the Camino ICP Monitor during use.
Problem
Cause
Recommendation
Camino ICP Monitor does not
operate with AC power.
AC power adapter not connected
Attach AC power adapter to Camino ICP Monitor.
Faulty AC power adapter cord
Attach AC adapter cord to Camino ICP Monitor.
•
Perform AC power test to ensure AC power works.
•
Contact Natus Neuro to order new AC power adapter cord.
Camino ICP Monitor does not
operate on battery power.
Battery not installed.
•
Attach AC power adapter to Camino ICP Monitor.
•
On touch screen, check battery indicator on status bar to
ensure the battery is installed.
•
Install battery.
Battery not charged
•
Attach AC power adapter to Camino ICP Monitor.
•
On touch screen, check battery indicator on status bar to
ensure the battery is installed.
Faulty battery
•
Attach AC adapter cord to Camino ICP Monitor.
•
Perform battery charge test to ensure battery holds charge
properly.
•
Contact Natus Neuro to order new battery.
Содержание Camino ICP
Страница 1: ...1 Camino ICP Monitor Intracranial Pressure and Temperature Monitor User s Manual 025870 Rev 00 01 2019...
Страница 2: ...2...
Страница 12: ...12 Blank page...
Страница 16: ...16 Blank page...
Страница 26: ...26 About the Left Panel Theleftpanel doesnotcontain anyusableconnector ports or buttons...
Страница 56: ...56 Blank page...
Страница 74: ...74 Blank page...
Страница 92: ...92 Blank page...
Страница 100: ...100 Blank page...
Страница 110: ...110 Blank page...
Страница 142: ...142 Blank page...
Страница 146: ...146 Blank page...