38
3.
After the setup process completes, the monitor will sound a one-second startup tone and display the
Main
panel:
Caution
The purpose of the startup tone verifies that the audio alarms are functioning correctly. If this tone does not sound
during the startup process, contact Natus Neuro for service.
Warning
To prevent injury to the patient, user, or other persons, or damage to the monitor, make sure to position the cables so that
they are free from all foot traffic.
Turning Off the System
On the front of the monitor, press the
Power
button. If the monitor ever freezes and does not turn off, press and hold
the
Power
button for several seconds to perform a forced shutdown of the system.
Using the Battery for Power
The Camino ICP Monitor includes a rechargeable 14.4 V lithium ion battery that supplies power to the monitor for
at least 1.5 hours when the battery is fully charged; the battery is only intended for use during patient transport.
•
To charge the battery to full capacity, turn off the monitor and plug it into an AC outlet for at least 5 hours prior to
use.
•
To ensure the battery maintains charge during patient use, always plug the monitor to an AC outlet whenever
possible.
If the battery exhibits problems powering the monitor for 1.5 hours, perform the "Battery Charge" test (Chapter 9)
to ensure the battery is functioning properly.
Содержание Camino ICP
Страница 1: ...1 Camino ICP Monitor Intracranial Pressure and Temperature Monitor User s Manual 025870 Rev 00 01 2019...
Страница 2: ...2...
Страница 12: ...12 Blank page...
Страница 16: ...16 Blank page...
Страница 26: ...26 About the Left Panel Theleftpanel doesnotcontain anyusableconnector ports or buttons...
Страница 56: ...56 Blank page...
Страница 74: ...74 Blank page...
Страница 92: ...92 Blank page...
Страница 100: ...100 Blank page...
Страница 110: ...110 Blank page...
Страница 142: ...142 Blank page...
Страница 146: ...146 Blank page...