Transient Pressure Changes During Implantation or Removal of Catheter May
Trigger Alarm
If the Flex Catheter is connected to the monitor during the implantation or removal of a catheter, a sudden
change in ICP measurements may occur while the catheter is traveling. Depending on the level of pressure
change, the ICP measurements may temporarily exceed the ICP accuracy range for the monitor (between -10
mmHg and 125 mmHg) and trigger one of the following alarms:
•
"
ICP out of accuracy range
";
or
•
"
ICP catheter failure
"
If either of these alarms is triggered during implantation or removal of the catheter, disconnect and reconnect the
Flex Catheter to the Flex Extension Cable.