Testing AC Power and Battery Charge
AC Power
Natus Neuro recommends performing this test once every three months:
1.
Plug the Natus Camino ICP Monitor to an AC power outlet.
2.
Press the
Power
button to turn on the monitor.
3.
Verify the following:
•
On the front of the monitor, check that the LED AC power indicator that appears above the power button is
illuminated.
•
On the touch screen, verify the plug icon appears in the status bar.
4.
On the rear of the monitor, disconnect the power cord and verify the following:
•
On the front of the monitor, check that the green LED AC power indicator turns off.
•
On the touch screen, verify the plug icon appears crossed-out in the status bar.
If you experience AC power problems, contact Natus Neuro.