120
Temperature Transducer Connector Test Circuit Diagram
Adjust the 10K
Ω
pot for the voltages at Pin 2 above
8.
Turn on the monitor, and adjust the 10K
Ω
potentiometer to produce a reading on the DMM of 0.350 VDC.
On the touch screen, the
Temperature
value should read 30.0 ± 0.3 °C.
9.
Adjust the
10KΩ
potentiometer to produce a reading on the DMM of 0.438 VDC. T On the touch screen, the
Temperature
value should read 40.0 ± 0.3 °C.
10.
Turn off monitor. On the right side of the monitor, remove all components from the
Temperature
connector.