System Message
Cause
Recommendation
Battery failure
Battery’s control system has failed.
Plug Camino ICP Monitor into AC power outlet.
After patient treatment, turn off Camino ICP Monitor and
then turn back on.
Replace the battery.
If problem persists, contact Natus Neuro.
Battery is low
Battery has 15 minutes or less of charge.
Plug Camino ICP Monitor into AC power outlet.
Battery has 5 minutes or less of charge
(accompanied by alarm tone).
Plug Camino ICP Monitor into AC power outlet.
Cannot write to log file
Cannot write data to log file that is used by
Natus Neuro for servicing.
Not resolvable. Contact Natus Neuro.