6.
Fill a water bath or beaker with warm water at approximately 32 °C and do the following:
A. Place the Fiber Optic Catheter in the water.
B. If necessary, place thermometer in water.
C. On the touch screen, press the
Main
tab and verify the
Temperature
value matches the
thermometer reading to ± .3 of a degree C.
7.
Using the table that appears after step 10, verify that the measured resistance is between the low and
high limits for the displayed temperature value.
8.
Fill a water bath or beaker with warm water at approximately
38
°C and do the following:
A. Place the Fiber Optic Catheter in the water.
B. On the touch screen, press the
Main
tab and verify the
Temperature
value match the
thermometer reading of ± .3 of a degree C.
9.
Using the table that appears after step 10, verify that the measured resistance is between the low and
high limits for the displayed temperature value.
10.
Disconnect DMM, and turn off monitor.