53
Step 4: Check Additional Values to Confirm Synchronization
Notice
The following synchronization checks at 25 mmHg, 50 mmHg, and 100 mmHg are not required, but are recommended by
Natus Neuro.
On the ICP Camino’s touch screen, press the
Check Additional Values
button to transmit additional ICP values to
the patient bedside monitor for scaling verification:
•
25 mmHg
•
50 mmHg
•
100 mmHg
Note that pressing
Check Additional Values
each time advances to the next ICP value. If you are not checking
additional values, proceed to the final step, "Complete the Synchronization Process" in Chapter 3
A. On the Camino ICP Monitor’s touch screen, press the
Check Additional Values
button to
transmit an ICP value at 25 mmHg for up to one minute to the patient bedside monitor.
A countdown will appear to display the amount time left to confirm that the value is also displayed on the patient
bedside monitor.
B. Once the ICP value on the patient bedside monitor stabilizes, verify that the patient bedside monitor
reads the same value as the Camino ICP Monitor:
25 mmHg, ± 1 mmHg
C. Repeat steps A to B in this section for each additional value that you want to check.
Содержание Camino ICP
Страница 1: ...1 Camino ICP Monitor Intracranial Pressure and Temperature Monitor User s Manual 025870 Rev 00 01 2019...
Страница 2: ...2...
Страница 12: ...12 Blank page...
Страница 16: ...16 Blank page...
Страница 26: ...26 About the Left Panel Theleftpanel doesnotcontain anyusableconnector ports or buttons...
Страница 56: ...56 Blank page...
Страница 74: ...74 Blank page...
Страница 92: ...92 Blank page...
Страница 100: ...100 Blank page...
Страница 110: ...110 Blank page...
Страница 142: ...142 Blank page...
Страница 146: ...146 Blank page...