A. On the rear panel of the Natus Neuro monitor, attach the connector end of the PMIO cable to the port
labeled
PMIO.
•
Align the red dot on the PMIO cable connector with the red triangle on the monitor’s port and push
firmly.
•
Verify the cable is completely connected to the monitor before proceeding.
1
A C power adapter cable
2
PMIO cable
B. On the other end of the PMIO cable, attach the monitor adapter cable for pressure (and temperature
if applicable).
C. Attach the PMIO cable and adapter cable to the temperature/pressure modules on the patient bedside
monitor.
Notice
The Camino ICP Monitor does not measure Cerebral Perfusion Pressure (CPP) ranges. As a result, the CPP
connector on the PMIO cable will not be used.
Step 2: Press the "Synchronize to Monitor" Button
On the Camino ICP Monitor’s touch screen, press the
Synchronize to Monitor
button; the Natus Neuro monitor
will transmit an ICP value of 0 mmHg for up to one minute. A countdown will appear to display the amount time left
to zero the patient bedside monitor.
Step 3: Zero the Patient Bedside Monitor to the Natus Neuro Monitor
On the patient bedside monitor, follow the manufacturer’s instructions to zero the monitor.
Содержание Camino ICP
Страница 1: ...1 Camino ICP Monitor Intracranial Pressure and Temperature Monitor User s Manual 025870 Rev 00 01 2019...
Страница 2: ...2...
Страница 12: ...12 Blank page...
Страница 16: ...16 Blank page...
Страница 26: ...26 About the Left Panel Theleftpanel doesnotcontain anyusableconnector ports or buttons...
Страница 56: ...56 Blank page...
Страница 74: ...74 Blank page...
Страница 92: ...92 Blank page...
Страница 100: ...100 Blank page...
Страница 110: ...110 Blank page...
Страница 142: ...142 Blank page...
Страница 146: ...146 Blank page...