54
Step 5: Complete the Synchronization Process
After completing the synchronization process, press
Done
. Once the ICP value on the patient bedside monitor
stabilizes, verify that the patient bedside monitor reads the same ICP value as the Camino ICP Monitor (±1 mmH
or 1 % of monitor reading, whichever is greater).
During patient treatment, Natus Neuro recommends comparing the ICP values between the two monitors while
the clinical staff is:
•
Recording ICP
•
Treating ICP
•
Per the hospital’s unit policy.
Warning
If the Camino ICP Monitor loses power and shuts down while it is connected to a patient bedside monitor, do not use the
ICP values on the patient bedside monitor for patient measurements; the ICP values on the patient bedside monitor will be
invalid.
Notice
After completing the synchronization process, always use the Camino ICP Monitor’s measurements over the
patient bedside monitor’s measurements. If drift occurs between the two monitors, use the Camino ICP
Monitor’s ICP values for the patient’s measurements and repeat the procedures for synchronizing the two
monitors (see Chapter 4).
About Pressure and Temperature Measurements on the Patient Bedside Monitor
The Camino ICP Monitor is designed to measure tissue pressure between the range of -10 mmHg to 125 mmHg, and
tissue temperature between the range of 30 °C and 42 °C. If the pressure or temperature value is outside these
specific ranges on the Camino ICP Monitor while it is connected to the patient bedside monitor, the Natus Neuro
monitor will activate a low priority alarm for either an ICP or temperature accuracy range alarm condition).
Depending on the type of alarm condition, the Natus Neuro monitor will also transmit either a pressure value of-
15 mmHg or a temperature value of 15 °C to the patient bedside monitor to indicate that the Natus Neuro monitor
is unable to make accurate measurements.
Содержание Camino ICP
Страница 1: ...1 Camino ICP Monitor Intracranial Pressure and Temperature Monitor User s Manual 025870 Rev 00 01 2019...
Страница 2: ...2...
Страница 12: ...12 Blank page...
Страница 16: ...16 Blank page...
Страница 26: ...26 About the Left Panel Theleftpanel doesnotcontain anyusableconnector ports or buttons...
Страница 56: ...56 Blank page...
Страница 74: ...74 Blank page...
Страница 92: ...92 Blank page...
Страница 100: ...100 Blank page...
Страница 110: ...110 Blank page...
Страница 142: ...142 Blank page...
Страница 146: ...146 Blank page...