43
A. Attach the cable’s ICP connector (
P
) into the port labeled
ICP (CAMCABL)
.
To attach:
•
Align the red dot on the cable connector with the red triangle on the monitor’s port and push firmly.
•
Verify the cable is completely connected to the monitor before proceeding.
B.
If temperature will be monitored for the patient, attach the cable’s temperature connector (
T
) into the
port labeled
ICT (CAMCABL)
.
To attach:
•
Align the red dot on the cable connector with the red triangle on the monitor’s port and push firmly.
•
Verify the cable is completely connected to the monitor before proceeding.
2. When operating in the sterile field, select the desired Camino Pressure Monitoring Kit (model 110-4 series).
For instructions on using the kit, see the directions for use supplied with each kit.
3.
Remove the catheter from its sterile tray and firmly attach its ICP and temperature connections to the Fiber
Optic Catheter Cable:
A. Attach the ICP transducer from the catheter to the ICP connector on the Fiber Optic Catheter Cable.
B. If applicable, attach the temperature thermistor from the catheter to the temperature connector on the
Fiber Optic Catheter Cable.
1
Temperature thermistor from catheter
2
Temperature connector from catheter
cable
3
ICP transducer from catheter
4
ICP connector from catheter cable
Содержание Camino ICP
Страница 1: ...1 Camino ICP Monitor Intracranial Pressure and Temperature Monitor User s Manual 025870 Rev 00 01 2019...
Страница 2: ...2...
Страница 12: ...12 Blank page...
Страница 16: ...16 Blank page...
Страница 26: ...26 About the Left Panel Theleftpanel doesnotcontain anyusableconnector ports or buttons...
Страница 56: ...56 Blank page...
Страница 74: ...74 Blank page...
Страница 92: ...92 Blank page...
Страница 100: ...100 Blank page...
Страница 110: ...110 Blank page...
Страница 142: ...142 Blank page...
Страница 146: ...146 Blank page...