80
Body
<VS-110>
10. Clean and Dismantle of Vacuum Syringe/Saliva Ejector
CAUTION
Do not conduct ultrasonic cleaning for disassembled parts of the adjustment
lever. Also, do not rub the surface of the parts with a sharp apparatus.
>>Surface of the parts can be damaged, and it would disturb to movement.
Once in a month or in case the movement of the suction adjustment part
goes sticky or hard, clean disassembled parts of the vacuum syringe.
(1)Detach the body from the main tube.
(2)Disassemble the body as a right illustration and take the drum off.
(3)Clean each part with running water.
Remove harden substances adhering to each part carefully with a tooth-
brush.
In the case of SE-110, check the state of the O-ring at
both ends of the drum. If there are transformation, hard-
ening and deficiency on the O-ring, replace it with a new
one.
(4)Reassemble the parts into an original form.
The lever will move smoothly if you apply a small quantity
of Vaseline is to the drum.
As for the VS-110, Insert the drum into the body so that
the grooves of the drum and the guide inside the body fit.
Then check the drum and the lever are linked together.
[6] Maintenance 10.
ADVICE
CHECK
CHECK
Drum
O-Ring
< SE-110 > < VS-110 >
Adjustment
lever
(Groove)