41
OFF
ON
Warmer Switch
Warmer Temperature Adjustment
Cold
@Hot
With a miniature flat head screw
driver
[5]Operation 7.
(2)Warmer
Whenever the warmer is turned on pay close attention to how hot the water
is. Turn the warmer off whenever it is not needed such as in the summer.
>>Otherwise, the water could get too hot and burn the patient. Also it could over-
heat and breakdown.
CAUTION
When the warmer is on, all the instruments and the auto cup filler receive
a supply of warm water.
1)Warmer Switch
The switch that turns the warmer on and off is located on the basin
unit panel.
When the switch is turned on, a light above it lights up and the warmer
starts to work. The instruments will receive warm water once the heater
has had enough time to warm it up.
2)Warmer Temperature Adjustment
The warmer temperature is set at the factory before shipping, but it
can be adjusted in the following way:
a)Turn the temperature adjustment to change the temperature of the
water.Use a miniature flat head screwdriver.
b)After waiting for a while, check the temperature of the cup water and
the threeway syringe. Do this several times if necessary.
When the warmer is on, the water will not be uni-
formly warm throughout the various systems.
Watch out for places that are hot and be aware of
the temperatures of various parts of the system.
c)Check that the water temperature is proper for use.
CHECK