69
Cut Filter
The switch
shown
length way
CAUTION
CAUTION
4. Cleaning the Basin Unit Component
[6]Maintenance 4.
Clean the items mentioned as described below at least once a day.
>>Otherwise, the equipment could malfunction, its performance could deteriorate,
or lines could become plugged.
(1)Inside of Main Tube for Vacuum Syringe
and Saliva Ejector
Use the procedure below to clean these tubes after finishing with each
patient:
1)Turn main switch on.
2)Take the vacuum syringe and saliva ejector off their holders on the
assistant’s side and suck up one or two cups of water.
Same time, take action while the valve lever in closing situation and
open the lever slowly to vacuum.
3)Return the instruments to their holders.
(2)Cleaning the Cut Filters
Make sure the Filter is properly and firmly in place when you replace after
cleaning.
>>Otherwise the vacuum syringe and saliva ejector will perform poorly, and there
is a risk of malfunctions or plugged lines.
Follow the procedure below to clean the cut filter (2 places) at least
once a day:
1)Take the cut filter off the unit.
Pick the cap head and then unscrew it slowly.
2)Separate the cap and the cut filter and then clean each with running
water.
3)Put the cut filter and the cap together and then put those ones back
to the unit.Insert the cap firmly and let each knob vertical.
Apply Vaseline to the O-ring at the side of the cap
so that the cap goes in and out smoothly.
ADVICE