40
Drain hole
[5]Operation 7.
Flushing Lamp
Flushing Switch
c)After connecting all the tubes, press the flushing switch.
The flushing procedure will be performed automatically.
It takes approximately 6 minutes.
Switch lamps will light up to show the progress.
<Stages>
Stage 1:Three-way water flows (continues to end of procedure)
Stage 2:Basin water flows (approximately 30 seconds)
Stage 3:Water flows from auto filler and vacuum and saliva ejector
are activated (approximately 30 seconds)
Stage 4:Water flows from instrument tubes
(approximately 30 seconds)
Stage 5:Repeat stages 3 and 4 above 4 times.
Stage 6:Vacuum and saliva ejector are activated
(approximately 15 seconds)
d)The lamp above the switch goes out when the procedure is finished.
Disconnect all the tubes and put them back in their holders.
Be sure to release the retainer holding down the
water lever on the three-way syringe. Something
could be damaged if the tube is pulled off without
doing this first.
e)Take the flushing bowl off the basin.
Most of the water inside the device is removed
by the vacuum and saliva ejector, but a little may
be left. If so, pour it out before putting the device
away.
If considerable water is left in the bowl, drain it
out through the drain hole.
f)Take Bowl Apart
The top and bottom of the bowl can be taken apart for cleaning.
Loosen the screw in the middle of the bowl.
Do not tighten the screw too much when putting
the bowl back together. This could damage it.
Screw
ADVICE
CHECK
ADVICE
Bowl set up : Lamp on
Flushing : Lamp Blinks
Take Bowl : Lamp Off
Flushing Lamp
Stage1 ~ 3,6: Amber Lamp Blinks
Stage4: Green Lamp Blinks
Finished: Amber Lamp Off
Spray Switch