29
(1)Air Turbine Hand piece
1)Operation
Take the handpiece out of its holder and step on the foot pedal to op-
erate it.
When the spray is on, rotation and spay can be operated by the foot
pedal.
2)Air Turbine variable speed settings*
In case the variable speed control is equipped optionally, rotation
between low and high can be adjusted steplessly. Also, conversion
(constant or variable)of rotation speed mode is possible.
When the spray is on, rotation and spay can be operated by the foot
pedal.
3)Clean System*
This prevents the vacuum effect when the hand piece stops running
and thereby prevents contamination being sucked into the system.
When the turbine stops,the IN side is closed and a little air flows out
of the OUT side for about 10 seconds.
[5]Operation 6.
*:option
(2)Air Scaler
*
An air scaler can be connected to the air turbine system.
It will then operate in the same way as an air turbine.
In case you use an air scaler, the optional equipment of
the air turbine with the variable speed control* is
recommended. The power of vibration is adjustable with
the variable speed control.
ADVICE
Splay ON(Variable*)
Pedal Rotaion Splay
Step Variable OFF-ON
Full Step Top speed ON
Splay ON(Basic)
Pedal Rotaion Splay
Step Hold OFF-ON
Full Step Hold ON